User Tools

Site Tools


venting_pumping_intermediate_image_box

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Next revision Both sides next revision
venting_pumping_intermediate_image_box [2024/01/11 10:33]
noji [Venting the Chamber]
venting_pumping_intermediate_image_box [2024/01/11 10:37]
noji [Venting the Chamber]
Line 32: Line 32:
   * that the PPACs have been unbiased,   * that the PPACs have been unbiased,
   * that this pumping station has been isolated by having the gate valves **D2426 BGV** and **D2436 BGV** closed, and that the IG has been turned off, and   * that this pumping station has been isolated by having the gate valves **D2426 BGV** and **D2436 BGV** closed, and that the IG has been turned off, and
-  * that the turbo pump gate valve TVG is closed.+  * that the turbo pump gate valve **TVG** is closed
 + 
 +The TMP can be turned off at this time.
  
 To vent the chamber: To vent the chamber:
-  - Put the GHS in vent mode by clicking **VENT**,+  - Put the GHS in vent mode by clicking **VENT** (valves #5 and #6 are open),
   - monitor **RBT_BTS32:BPG_D2430:VP_RD** and **S800_GHS:PI_D2431:V_RD_AB** in a data browser,   - monitor **RBT_BTS32:BPG_D2430:VP_RD** and **S800_GHS:PI_D2431:V_RD_AB** in a data browser,
   - close **RV**, the valve to the roughing pump, and   - close **RV**, the valve to the roughing pump, and
venting_pumping_intermediate_image_box.txt · Last modified: 2024/01/12 09:37 by noji