Line 22: Line 22:
  
 ===== Emptying Detectors ===== ===== Emptying Detectors =====
 +  - Double check that the detectors have been unbiased.
   - Set “PPAC Setpoint” to 0 Torr.   - Set “PPAC Setpoint” to 0 Torr.
   - Click **EMPTY** on the “GHS_II-PPAC_Main” page to open valves #3 and #4 which will allow for pumping out the gas quickly.   - Click **EMPTY** on the “GHS_II-PPAC_Main” page to open valves #3 and #4 which will allow for pumping out the gas quickly.
Line 37: Line 38:
  
 To vent the chamber: To vent the chamber:
-  - Put the GHS in vent mode by clicking **VENT** (valves #5 and #6 are open),+  - Let the Control Room and the Vacuum Operation Group be aware as EXP alarms may occur, 
 +  - put the GHS in vent mode by clicking **VENT** (valves #5 and #6 are open),
   - monitor **RBT_BTS32:BPG_D2430:VP_RD** and **S800_GHS:PI_D2431:V_RD_AB** in a data browser,   - monitor **RBT_BTS32:BPG_D2430:VP_RD** and **S800_GHS:PI_D2431:V_RD_AB** in a data browser,
-  - close **RV**, the valve to the roughing pump, and+  - ensure that **RV**, the valve to the roughing pump, is closed, and
   - open **Vent** valve to start venting.   - open **Vent** valve to start venting.
  