Line 37: Line 37:
  
 To vent the chamber: To vent the chamber:
-  - Put the GHS in vent mode by clicking **VENT** (valves #5 and #6 are open),+  - Let the Control Room and the Vacuum Operation Group be aware as EXP alarms may occur, 
 +  - put the GHS in vent mode by clicking **VENT** (valves #5 and #6 are open),
   - monitor **RBT_BTS32:BPG_D2430:VP_RD** and **S800_GHS:PI_D2431:V_RD_AB** in a data browser,   - monitor **RBT_BTS32:BPG_D2430:VP_RD** and **S800_GHS:PI_D2431:V_RD_AB** in a data browser,
   - ensure that **RV**, the valve to the roughing pump, is closed, and   - ensure that **RV**, the valve to the roughing pump, is closed, and