User Tools

Site Tools


venting_pumping_object_box

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision Both sides next revision
venting_pumping_object_box [2023/05/30 14:31]
swartzj
venting_pumping_object_box [2023/05/30 14:32]
swartzj
Line 11: Line 11:
   - Close the valve to the roughing pump RV.    - Close the valve to the roughing pump RV. 
   - Now you can plot the pressure at BPG to monitor the pressure before you open the vent valve.    - Now you can plot the pressure at BPG to monitor the pressure before you open the vent valve. 
-  - **Keep a constant eye on the pressure level** as it increases rapidly. Before it nears 800 torr, **it is imperative that you close the venting valve.** Failure to do this will result in over pressure since the chamber will continue to be filled with dry nitrogen.  +  - **Keep a constant eye on the pressure level** as it increases rapidly. Before it nears 800 torr, **it is imperative that you close the venting valve.** Failure to do so will result in over pressure since the chamber will continue to be filled with dry nitrogen.  
   - You can now enter and open the chamber.    - You can now enter and open the chamber. 
  