venting_pumping_object_box [2023/05/30 14:32] swartzj |
venting_pumping_object_box [2023/05/30 16:48] swartzj |
||
---|---|---|---|
Line 6: | Line 6: | ||
* Before you start, make sure that the gate valves named D2341 BGV and D2303 BGV are closed and that the IG is off. | * Before you start, make sure that the gate valves named D2341 BGV and D2303 BGV are closed and that the IG is off. | ||
* The vent valve should also be closed when you begin. | * The vent valve should also be closed when you begin. | ||
- | * Both valves to the turbo pump should be closed and the turbo pump should be switched off. | + | * Both valves to the turbo pump should be closed and the turbo pump should be switched off if the chamber will be vented for a long time. |
The procedure is then as follows: | The procedure is then as follows: | ||
- Close the valve to the roughing pump RV. | - Close the valve to the roughing pump RV. | ||
- Now you can plot the pressure at BPG to monitor the pressure before you open the vent valve. | - Now you can plot the pressure at BPG to monitor the pressure before you open the vent valve. | ||
- | - **Keep a constant eye on the pressure level** as it increases rapidly. | + | - **Keep a constant eye on the pressure level** as it increases rapidly. |
- You can now enter and open the chamber. | - You can now enter and open the chamber. | ||