This shows you the differences between two versions of the page.
Both sides previous revision Previous revision | Next revision Both sides next revision | ||
venting_pumping_object_box [2023/06/01 15:50] swartzj |
venting_pumping_object_box [2023/06/02 17:13] swartzj |
||
---|---|---|---|
Line 19: | Line 19: | ||
* Make sure that the IG is off. | * Make sure that the IG is off. | ||
* When you start, you should have RV closed, FV open and the TGV valve, Vent valve, as well as the gate valves named D2341 BGV and D2303 BGV all closed. | * When you start, you should have RV closed, FV open and the TGV valve, Vent valve, as well as the gate valves named D2341 BGV and D2303 BGV all closed. | ||
- | * Close FV and open RV to start pumping down the object chamber with the roughing pump. You can make a plot of the pressure at BPG at this point. | + | * Close FV and open RV to start pumping down the object chamber with the roughing pump. You can make a plot of the pressure |
* Section D2337 will initially be red with vacuum status: 'Bad vacuum' | * Section D2337 will initially be red with vacuum status: 'Bad vacuum' | ||
* Wait until the vacuum reading at FPG to drops to below 200 mTorr before you can safely turn on the turbo pump. | * Wait until the vacuum reading at FPG to drops to below 200 mTorr before you can safely turn on the turbo pump. |