Line 33: Line 33:
  
  
-Each detector consists of two windows mounted on frames, two printed circuit boards (PCB) and an anode frame. Each PCB is made of un-masked G-10, and includes a field shaping foil to ensure a uniform field in the active region of the detector. Two G-10 spacers are laminated to the board on each side. The shaping foils are made of 1.9-mm pitch evaporated aluminum strips perpendicularly oriented to the electric field. The anode frame includes a glued cathode grounding plane, an anode wire running across the field, and a Frisch grid. Cathode pads are located in front of and behind the anode wire. The pads have a pitch of 2.54 mm. The anode frame is sandwiched between the two printed circuit boards with two spacers in between, as shown in the figure below.+Each detector consists of two 12-µm PPTA windows mounted on frames, two printed circuit boards (PCB) and an anode frame. Each PCB is made of un-masked G-10, and includes a field shaping foil (70 µg/cm<sup>2</sup> polypropylene with 0.1 µm of evaporated gold) to ensure a uniform field in the active region of the detector. Two G-10 spacers are laminated to the board on each side. The shaping foils are made of 1.9-mm pitch evaporated aluminum strips perpendicularly oriented to the electric field. The anode frame includes a glued cathode grounding plane, an anode wire running across the field, and a Frisch grid. Cathode pads are located in front of and behind the anode wire. The pads have a pitch of 2.54 mm. The anode frame is sandwiched between the two printed circuit boards with two spacers in between, as shown in the figure below.
  
 {{:wiki:crdc-section.jpg?600|Cross section of a CRDCs.}} {{:wiki:crdc-section.jpg?600|Cross section of a CRDCs.}}
Line 52: Line 52:
  
 ===== Ionization chamber ===== ===== Ionization chamber =====
-An ionization chamber downstream of both [[Detectors#Cathode Readout Drift Chambers (CRDC)|CRDCs]] is used to identify the Z number of the transmitted nuclei from their energy loss. The detector has an active area of __xxx cm x xxx cm__ and a depth of approximately 406 mm (16 inches). It [[Gas handling system||is filled]] with P10 gas (90% argon, 10% methane) at a typical pressure of 300 torr, although this value can be increased up to 600 torr for light nuclei. The detector consists of 16 stacked-parallel plate ion chambers with narrow anode-cathode gaps, placed along the detector’s central axis, perpendicular to the beam direction (see figure). The plates are constructed from 70 mg/cm<sup>2</sup> polypropylene with 0.05 µm of aluminum evaporated on each side. The entrance and exit windows of the chamber are made of 14 mg/cm<sup>2</sup> Mylar with an overlay of Kevlar filaments and epoxy.+An ionization chamber downstream of both [[Detectors#Cathode Readout Drift Chambers (CRDC)|CRDCs]] is used to identify the Z number of the transmitted nuclei from their energy loss. The detector has an active area of approximately 30 cm x 60 cm and a depth of approximately 406 mm (16 inches). It [[Gas handling system|is filled]] with P10 gas (90% argon, 10% methane) at a typical pressure of 300 torr, although this value can be increased up to 600 torr for light nuclei. The detector consists of 16 stacked-parallel plate ion chambers with narrow anode-cathode gaps, placed along the detector’s central axis, perpendicular to the beam direction (see figure). The plates of each of these stacked chambers are constructed from 70 µg/cm<sup>2</sup> polypropylene with 0.15 µm of aluminum evaporated on each side. The entrance and exit windows of the chamber are made of 12 mg/cm<sup>2</sup> Mylar with an overlay of Kevlar filaments and epoxy.
 {{:wiki:ion-chamber-picture.jpg?500 |Picture of the S800 ionization chamber with its alternating cathode and anode plates.}} {{:wiki:ion-chamber-picture.jpg?500 |Picture of the S800 ionization chamber with its alternating cathode and anode plates.}}
-The electrons and positive ions liberated by the ionization of the gas along the particle trajectory drift towards the closest  anode-cathode pair. The drifting electrons and ions absorb the energy stored in the detector capacity and produce a voltage change of the anodes across the resistor. The main advantages of the anode-cathode configuration is that the electrons and ions are collected on a very short distance (about 1.5 cm), thus reducing pile-up and position dependence of the signals. Moreover, dividing the detector into 16 sections reduces the detector capacitance and consequently its noise. The operating voltage depends on the charge of the measured nuclei (e.g. __xxx for xxx and xxx for xxx__). Each anode is attached to a small preamplifier inside the ion chamber. This significantly reduces the electronic noise, although it involves the venting of the whole chamber whenever a malfunctioning preamplifier needs to be replaced. The electronic signals from the preamplifier are sent into a [[https://groups.nscl.msu.edu/nscl_library/manuals/caen/MOD.N568B.pdf|CAEN N568B]] 16-channel shaper/amplifier with remotely adjustable gains. The output signals feed a [[https://groups.nscl.msu.edu/nscl_library/manuals/phillips/7164H.pdf|Phillips 7164H]] ADC. + 
 +The electrons and positive ions liberated by the ionization of the gas along the particle trajectory drift towards the closest  anode-cathode pair. The drifting electrons and ions absorb the energy stored in the detector capacity and produce a voltage change of the anodes across the resistor. The main advantages of the anode-cathode configuration is that the electrons and ions are collected on a very short distance (about 1.5 cm), thus reducing pile-up and position dependence of the signals. Moreover, dividing the detector into 16 sections reduces the detector capacitance and consequently its noise. Each anode is attached to a small preamplifier inside the ion chamber. This significantly reduces the electronic noise, although it involves the venting of the whole chamber whenever a malfunctioning preamplifier needs to be replaced. The electronic signals from the preamplifier are sent into a [[https://groups.nscl.msu.edu/nscl_library/manuals/caen/MOD.N568B.pdf|CAEN N568B]] 16-channel shaper/amplifier with remotely adjustable gains. The output signals feed a [[https://groups.nscl.msu.edu/nscl_library/manuals/phillips/7164H.pdf|Phillips 7164H]] ADC. 
 The energy-loss resolution of the ionization chamber can be significantly improved after correcting the position and momentum dependences. Elements up to Z=50 can be separated. The energy-loss resolution of the ionization chamber can be significantly improved after correcting the position and momentum dependences. Elements up to Z=50 can be separated.
  