====== During experiment ====== | ====== During experiment ====== | ||
This section describes support tasks that are typically requested during the course of an experiment. Some of them, like Mask calibration, | This section describes support tasks that are typically requested during the course of an experiment. Some of them, like Mask calibration, | ||
+ | |||
+ | * [[#Mask calibration]] | ||
+ | * [[#Particle identification corrections]] | ||
+ | * [[# | ||
+ | |||
Line 90: | Line 95: | ||
- | ===== Handling detectors | + | ===== Efficiency of OBJ Scintillator |
+ | At relatively high rates (~0.5 MHz), the OBJ scintillator will be gradually damaged. Experimenters should check the efficiency of this detector continuously during the course of the experiment. Whenever the rates drop below ~90%, they may request the Device/Beam Physicist to shim the detector. | ||
+ | |||
+ | |||
+ | |||
+ | Before shimming, make sure that the thresholds are set properly (not too high) (refer to section " | ||