====== During experiment ====== | ====== During experiment ====== | ||
This section describes support tasks that are typically requested during the course of an experiment. Some of them, like Mask calibration, | This section describes support tasks that are typically requested during the course of an experiment. Some of them, like Mask calibration, | ||
+ | |||
+ | * [[#Mask calibration]] | ||
+ | * [[#Particle identification corrections]] | ||
+ | * [[# | ||
+ | |||
Line 18: | Line 23: | ||
* Set the trigger to S800 singles and downscale factor 1 | * Set the trigger to S800 singles and downscale factor 1 | ||
- | * On PanelMate, open page 4 in S800DRIV | + | |
- | * Click on "upstream mask" | + | |
+ | |||
+ | | ||
* Start a run without taking to disk and look in the S800 SpecTcl at the 2D xy position spectrum of (s800.fp.crdc1.x, | * Start a run without taking to disk and look in the S800 SpecTcl at the 2D xy position spectrum of (s800.fp.crdc1.x, | ||
Line 90: | Line 97: | ||
- | ===== Handling detectors | + | ===== Efficiency of OBJ Scintillator |
+ | At relatively high rates (~0.5 MHz), the OBJ scintillator will be gradually damaged. Experimenters should check the efficiency of this detector continuously during the course of the experiment. Whenever the rates drop below ~90%, they may request the Device/Beam Physicist to shim the detector. | ||
+ | |||
+ | |||
+ | |||
+ | Before shimming, make sure that the thresholds are set properly (not too high) (refer to section " | ||
==== Shimming OBJ scintillator ==== | ==== Shimming OBJ scintillator ==== |