This shows you the differences between two versions of the page.
Both sides previous revision Previous revision Next revision | Previous revision Last revision Both sides next revision | ||
during_experiments [2017/06/17 14:46] pereira [During experiment] |
during_experiments [2023/07/20 10:41] swartzj |
||
---|---|---|---|
Line 4: | Line 4: | ||
* [[#Mask calibration]] | * [[#Mask calibration]] | ||
* [[#Particle identification corrections]] | * [[#Particle identification corrections]] | ||
- | * [[#Handling detectors|Efficiency of OBJ scintillator]] | + | * [[# |
Line 23: | Line 23: | ||
* Set the trigger to S800 singles and downscale factor 1 | * Set the trigger to S800 singles and downscale factor 1 | ||
- | * On PanelMate, open page 4 in S800DRIV | + | |
- | * Click on "upstream mask" | + | |
+ | |||
+ | | ||
* Start a run without taking to disk and look in the S800 SpecTcl at the 2D xy position spectrum of (s800.fp.crdc1.x, | * Start a run without taking to disk and look in the S800 SpecTcl at the 2D xy position spectrum of (s800.fp.crdc1.x, | ||
Line 95: | Line 97: | ||
- | ===== Handling detectors | + | ===== Efficiency of OBJ Scintillator |
+ | At relatively high rates (~0.5 MHz), the OBJ scintillator will be gradually damaged. Experimenters should check the efficiency of this detector continuously during the course of the experiment. Whenever the rates drop below ~90%, they may request the Device/Beam Physicist to shim the detector. | ||
+ | |||
+ | |||
+ | |||
+ | Before shimming, make sure that the thresholds are set properly (not too high) (refer to section " | ||
==== Shimming OBJ scintillator ==== | ==== Shimming OBJ scintillator ==== |