Line 96: Line 96:
  
 ===== Efficiency of OBJ Scintillator ===== ===== Efficiency of OBJ Scintillator =====
 +At relatively high rates (~0.5 MHz), the OBJ scintillator will be gradually damaged. Experimenters should check the efficiency of this detector continuously during the course of the experiment. Whenever the rates drop below ~90%, they may request the Device/Beam Physicist to shim the detector. 
 +
 +
 +
 +Before shimming, make sure that the thresholds are set properly (not too high) (refer to section "Object Scintillator Setup during [[Focus Mode#Tuning the S800 (XDT)|XDT]]" 
 +
  
 ==== Shimming OBJ scintillator ==== ==== Shimming OBJ scintillator ====