Line 15: Line 15:
   * PanelMate **S800vac.ui page 07**: Focal Plane Detector   * PanelMate **S800vac.ui page 07**: Focal Plane Detector
  
-The gas handling systems for both Focal Plane gas-filled detectors – the [[Detectors#Ionization Chamber|Ionization Chamber]] and the [[Detectors#Cathode Readout Drift Chambers (CRDC)|CRDCs]] – are remotely controlled via a single LabView application running on the computer [[Software|S800FP-PC]] in the S3 vault. This computer can be accessed remotely from the desktop of computer [[Software#u6pc5 (data U6)|u6pc5]], by clicking in the icon **S800 Vault PC**, or by typing ''rdesktop -d NSCL_USER -u s800 -g 1600x1000 35.9.56.104'' from a Linux terminal. Once the virtual session in S800FP-PC is open, the gas handling system can be opened by clicking in the corresponding icon on the desktop **FPGHS_TCPIP-II**. This icon opens the local directory ''C:/Users/s800/Desktop/FPGHS_TCPIP-II''. Once there, run the Labview executable **FPGHS_TCPIP-II.exe** +The gas handling systems for both Focal Plane gas-filled detectors – the [[Detectors#Ionization Chamber|Ionization Chamber]] and the [[Detectors#Cathode Readout Drift Chambers (CRDC)|CRDCs]] – are remotely controlled via a single LabView application running on the computer [[Software|S800FP-PC]] in the S3 vault. This computer can be accessed remotely from the desktop of computer [[Software#u6pc5 (data U6)|u6pc5]], by clicking in the icon **S800 Vault PC**, or by typing ''xfreerdp /u:s800 /v:35.9.56.104'' from a Linux terminal. A terminal will pop-up asking for the s800 password. Once the virtual session in S800FP-PC is open, the gas handling system can be opened by clicking in the corresponding icon on the desktop **FPGHS_TCPIP-II**. This icon opens the local directory ''C:/Users/s800/Desktop/FPGHS_TCPIP-II''. Once there, run the Labview executable **FPGHS_TCPIP-II.exe**
  
 Note that since the same roughing pump drives the gas handling systems for both gas-filled detectors, extra precautions must be taken to protect one detector from the pressure of the other detector. The fact that the same pump drives both systems is not indicated on the graphics for the LabView control application Note that since the same roughing pump drives the gas handling systems for both gas-filled detectors, extra precautions must be taken to protect one detector from the pressure of the other detector. The fact that the same pump drives both systems is not indicated on the graphics for the LabView control application
Line 42: Line 42:
       * Ensure **I265GV** is closed (in PanelMate **S800vac.ui page 07**) to isolate focal plane vacuum chamber from the rest of the beamline       * Ensure **I265GV** is closed (in PanelMate **S800vac.ui page 07**) to isolate focal plane vacuum chamber from the rest of the beamline
       * Confirm that the focal plane chamber is under vacuum       * Confirm that the focal plane chamber is under vacuum
-      * Confirm that the only open gas handling system valves are **4, 5, 6, 20, 21, and 22**+      * Confirm that the only open gas handling system valves are **4, 5, 6, 7, 20, 21, 22 and 23**
       * (Check section on [[vacuum control|vacuum control]] for further details.)       * (Check section on [[vacuum control|vacuum control]] for further details.)
  
Line 50: Line 50:
            * The display unit for the pressure at the pump has a needle indicator and is mounted under the [[Stations#Focal Plane Station|focal plane]] on the south support post             * The display unit for the pressure at the pump has a needle indicator and is mounted under the [[Stations#Focal Plane Station|focal plane]] on the south support post 
       * Ensure that the main valves on the gas bottles are open (see pictures below)       * Ensure that the main valves on the gas bottles are open (see pictures below)
 +      * Open valves downstream of regulators in each bottle
       * Ensure that the gas bottles are not empty       * Ensure that the gas bottles are not empty
            * The gas regulators are not normally adjusted. They should all be set to provide a pressure around 20 psi            * The gas regulators are not normally adjusted. They should all be set to provide a pressure around 20 psi
Line 71: Line 72:
           * Click the **Valves** tab at the top of the LabView window to return to the overview display of the gas handling system           * Click the **Valves** tab at the top of the LabView window to return to the overview display of the gas handling system
  
-  * Close **valve 6**+  * Close **valves 6** and **7**
  
   * Open **valve 2**; the flow reading at MFC #1 should jump momentarily   * Open **valve 2**; the flow reading at MFC #1 should jump momentarily
Line 102: Line 103:
           * Click the **Valves** tab at the top of the LabView window to return to the overview display of the gas handling system           * Click the **Valves** tab at the top of the LabView window to return to the overview display of the gas handling system
  
-  * Close **valve 22**+  * Close **valves 22** and **23**
    
   * Open **valves 10 and 15**   * Open **valves 10 and 15**
Line 156: Line 157:
   * Close **valves 8 and 24**   * Close **valves 8 and 24**
  
-  * Open **valves 6 and 22**+  * Open **valves 6, 7, 22 and 23**
  
-  * Enter S3 vault and close valves downstream of regulators in [[#Preparation of gas bottles|gas bottles]]: P-10 for the [[Detectors#Ionization Chamber|ionization chamber]] (C<sub>4</sub>H<sub>10</sub>) and CF<sub>4</sub> (tetrafluoromethane) for the [[Detectors#Cathode Readout Drift Chambers (CRDC)|CRDCs]] +  * Enter S3 vault and close valves downstream of regulators in [[#Preparation of gas bottles|gas bottles]]: P-10 and CF<sub>4</sub> (tetrafluoromethane). Close the main valve for the C<sub>4</sub>H<sub>10</sub> bottle.
  
 ===== Tracking Gas bottles pressures ===== ===== Tracking Gas bottles pressures =====