gas_handling_system [2018/04/09 12:07]
pereira [Filling CRDC1 and CRDC2 with Gas]
gas_handling_system [2018/04/11 16:12]
pereira [Filling CRDC1 and CRDC2 with Gas]
Line 72: Line 72:
           * Click the **Valves** tab at the top of the LabView window to return to the overview display of the gas handling system           * Click the **Valves** tab at the top of the LabView window to return to the overview display of the gas handling system
  
-  * Close **valves 6 and 7**+  * Close **valves 6** and **7**
  
   * Open **valve 2**; the flow reading at MFC #1 should jump momentarily   * Open **valve 2**; the flow reading at MFC #1 should jump momentarily
Line 103: Line 103:
           * Click the **Valves** tab at the top of the LabView window to return to the overview display of the gas handling system           * Click the **Valves** tab at the top of the LabView window to return to the overview display of the gas handling system
  
-  * Close **valve 22 and 23**+  * Close **valves 22** and **23**
    
   * Open **valves 10 and 15**   * Open **valves 10 and 15**
Line 157: Line 157:
   * Close **valves 8 and 24**   * Close **valves 8 and 24**
  
-  * Open **valves 6, 7, 22 and 23**+  * Open **valves 6 and 22**
  
   * Enter S3 vault and close valves downstream of regulators in [[#Preparation of gas bottles|gas bottles]]: P-10 and CF<sub>4</sub> (tetrafluoromethane). Close the main valve for the C<sub>4</sub>H<sub>10</sub> bottle.   * Enter S3 vault and close valves downstream of regulators in [[#Preparation of gas bottles|gas bottles]]: P-10 and CF<sub>4</sub> (tetrafluoromethane). Close the main valve for the C<sub>4</sub>H<sub>10</sub> bottle.