Line 10: Line 10:
   * [[#Replacing Gas bottles]]   * [[#Replacing Gas bottles]]
  
-===== General Information ===== +===== General Information for focal-plane gas handlying system =====
-Required controls and monitoring +
-  * LabView gas handling system application **FPGHS_TCPIP-II.vi**  +
-  * PanelMate **S800vac.ui page 07**: Focal Plane Detector +
 The gas handling systems for both Focal Plane gas-filled detectors – the [[Detectors#Ionization Chamber|Ionization Chamber]] and the [[Detectors#Cathode Readout Drift Chambers (CRDC)|CRDCs]] – are remotely controlled via a single LabView application running on the computer [[Software|S800FP-PC]] in the S3 vault. This computer can be accessed remotely from the desktop of computer [[Software#u6pc5 (data U6)|u6pc5]], by clicking in the icon **S800 Vault PC**, or by typing ''xfreerdp /u:s800 /v:35.9.56.104'' from a Linux terminal. A terminal will pop-up asking for the s800 password. Once the virtual session in S800FP-PC is open, the gas handling system can be opened by clicking in the corresponding icon on the desktop **FPGHS_TCPIP-II**. This icon opens the local directory ''C:/Users/s800/Desktop/FPGHS_TCPIP-II''. Once there, run the Labview executable **FPGHS_TCPIP-II.exe**.  The gas handling systems for both Focal Plane gas-filled detectors – the [[Detectors#Ionization Chamber|Ionization Chamber]] and the [[Detectors#Cathode Readout Drift Chambers (CRDC)|CRDCs]] – are remotely controlled via a single LabView application running on the computer [[Software|S800FP-PC]] in the S3 vault. This computer can be accessed remotely from the desktop of computer [[Software#u6pc5 (data U6)|u6pc5]], by clicking in the icon **S800 Vault PC**, or by typing ''xfreerdp /u:s800 /v:35.9.56.104'' from a Linux terminal. A terminal will pop-up asking for the s800 password. Once the virtual session in S800FP-PC is open, the gas handling system can be opened by clicking in the corresponding icon on the desktop **FPGHS_TCPIP-II**. This icon opens the local directory ''C:/Users/s800/Desktop/FPGHS_TCPIP-II''. Once there, run the Labview executable **FPGHS_TCPIP-II.exe**. 
  
 Note that since the same roughing pump drives the gas handling systems for both gas-filled detectors, extra precautions must be taken to protect one detector from the pressure of the other detector. The fact that the same pump drives both systems is not indicated on the graphics for the LabView control application Note that since the same roughing pump drives the gas handling systems for both gas-filled detectors, extra precautions must be taken to protect one detector from the pressure of the other detector. The fact that the same pump drives both systems is not indicated on the graphics for the LabView control application
 +
 +Required controls and monitoring
 +  * LabView gas handling system application **FPGHS_TCPIP-II.vi** 
 +  * PanelMate **S800vac.ui page 07**: Focal Plane Detector
  
 ===== LabView Control Program ===== ===== LabView Control Program =====
Line 142: Line 142:
   * Open **valves 5 and 21**   * Open **valves 5 and 21**
  
-  * **IMPORTANT: Before opening valve 8 (next step), ** CLOSE **VALVE 25** to protect CRDCs from Ion Chamber pressure+  * **IMPORTANT**: Before opening valve 8 (next step), CLOSE **VALVE 25** to protect CRDCs from Ion Chamber pressure
  
   * Empty Ion Chamber to a pressure at or below the pressure of the CRDCs   * Empty Ion Chamber to a pressure at or below the pressure of the CRDCs