Line 72: Line 72:
           * Click the **Valves** tab at the top of the LabView window to return to the overview display of the gas handling system           * Click the **Valves** tab at the top of the LabView window to return to the overview display of the gas handling system
  
-  * Close **valves 6 and 7**+  * Close **valves 6** and **7**
  
   * Open **valve 2**; the flow reading at MFC #1 should jump momentarily   * Open **valve 2**; the flow reading at MFC #1 should jump momentarily
Line 103: Line 103:
           * Click the **Valves** tab at the top of the LabView window to return to the overview display of the gas handling system           * Click the **Valves** tab at the top of the LabView window to return to the overview display of the gas handling system
  
-  * Close **valve 22**+  * Close **valves ​22** and **23**
    
   * Open **valves 10 and 15**   * Open **valves 10 and 15**
Line 142: Line 142:
   * Open **valves 5 and 21**   * Open **valves 5 and 21**
  
-  * **IMPORTANT:** CLOSE **VALVE 25** to protect CRDCs from Ion Chamber pressure+  * **IMPORTANT**: Before opening valve 8 (next step), ​CLOSE **VALVE 25** to protect CRDCs from Ion Chamber pressure
  
   * Empty Ion Chamber to a pressure at or below the pressure of the CRDCs   * Empty Ion Chamber to a pressure at or below the pressure of the CRDCs