====== Detector bias control ====== | ====== Detector bias control ====== | ||
+ | ===== HV Remote control ===== | ||
- | ===== General Information ===== | + | The high-voltage of the plastic scintillator and the MCP at Object, the [[Detectors# |
- | The [[Detectors# | + | |
+ | {{: | ||
- | {{: | + | This page can be reached from “FRIB EXP” → “DIAG” (Engineering Area, not to be confused with “S800” → “Diag Main”) → “S800 HV Bias Power Supply”. |
- | ===== Object Scintillator ===== | + | |
- | The bias for the object scintillator is controlled via the H.V. power supply {{: | + | |
- | ===== FP Scintillators ===== | + | The ramp rate can be changed |
- | Only the fist two scintillators are typically used since in most cases particles do not reach beyond | + | |
- | ===== Ion Chamber ===== | + | Each channel can also be accessed from by going to “FRIB EXP” -> “Power Supplies” |
- | The biases for the anode and the drift electrode are each controlled separately via the S800 Detector HV Control. Parts of this detector rely on the same 12-Volt power supply used to power the bases of the FP Scintillator PMTs. If the vacuum-based interlock condition for protecting the FP Scintillator PMTs is triggered, the 12-Volt power supply will not be available for the Ion Chamber. | + | |
- | ===== CRDCs ===== | + | {{: |
- | Each of the two detectors, CRDC1 and CRDC2, has a separate bias control for the anode and the drift electrode via the S800 Detector HV Control. Please keep in mind that a change in the drift voltage of the CRDCs strongly affects the drift velocity of the electrons and a new mask calibration has to be done to be able to determine the y position of particles. Apart from the CRDCs, it is important to bias the electrostatic gate (see section [[Detectors# | + | |
- | ===== TPPACs ===== | + | |
- | The bias of each of the two detectors, PPAC1 and PPAC2 is controlled individually via the S800 Detector | + | |
+ | The actual HV modules are located on the top level of the S3 vault, under the FP chamber. | ||
+ | {{: | ||
+ | |||
+ | Other examples of HV values used with different fragment beams are: | ||
+ | |||
+ | * **CRDC1 and CRDC2**: | ||
+ | * For He-3 @ ~130 MeV/u: CRDC1 (Anode=1120 V, Drift=1000 V); CRDC2 (Anode=1120 V; Drift=1000 V) | ||
+ | * For Be-12 @ ~30 MeV/u: CRDC1 (Anode=970 V, Drift=1000 V); CRDC2 (Anode=970 V; Drift=1000 V) | ||
+ | * For Ar-34 @ ~80 MeV/u: CRDC1 (Anode=860 V, Drift=1000 V); CRDC2 (Anode=840 V; Drift=1000 V) | ||
+ | * For Zn-58 @ ~44 MeV/u: CRDC1 (Anode=740 V, Drift=1000 V); CRDC2 (Anode=740 V; Drift=1000 V) | ||
+ | * For Rb-74 @ ~40 MeV/u: CRDC1 (Anode=650 V, Drift=1000 V); CRDC2 (Anode=650 V; Drift=1000 V) | ||
+ | * For U-238 (88+) @ ~70 MeV/u: CRDC1 (Anode=570 V, Drift=1000 V); CRDC2 (Anode=570 V; Drift=1000 V) | ||
+ | | ||
+ | * **FP scintillator (E1 up and E1 down)**: | ||
+ | * For He-3 @ ~130 MeV/u: UP (1790 V); DOWN (1760 V) | ||
+ | * For Be-12 @ ~30 MeV/u: UP (1770 V); DOWN (1700 V) | ||
+ | * For Ar-34 @ ~80 MeV/u: UP (1580 V); DOWN (1510 V) | ||
+ | * For Zn-58 @ ~44 MeV/u: UP (1400 V); DOWN (1340 V) | ||
+ | * For U-238 (88+) @ ~70 MeV/u: UP (1180 V); DOWN (1280 V) | ||
+ | |||
+ | * **IC**: | ||
+ | * Typically 200 V | ||