User Tools

Site Tools


hv_bias

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Next revision Both sides next revision
hv_bias [2014/11/19 11:59]
pereira [HV Manual control]
hv_bias [2023/02/13 14:29]
noji
Line 3: Line 3:
  
 ===== HV Remote control ===== ===== HV Remote control =====
-The high-voltage of the [[Detectors#Tracking Parallel Plate Avalanche Counters (TPPAC)|TPPACs]], [[Detectors#Ionization Chamber|Ion Chamber]], the [[Detectors#Cathode Readout Drift Chambers (CRDC)|CRDCs]], [[Detectors#Plastic scintillators|FP Scintillator]], and [[Detectors#Hodoscope|Hodoscope]] are remotely controlled via GUI control panel (see figure below). The GUI can be opened from the desktop of the u6pc5 control computer. This application allows also to set a maximum limit in the detector current. The voltages of these detectors depend on the Z of the fragments to be measured and their energy. The figure below shows the HV GUI during experiment e10002, which used Si-26 and Al-25 fragments at around 115 MeV/u. These voltages can be used as a reference before starting an experiment. However, the final values will have to be modified during the beam tuning in order to optimize the performance of the detectors. It is convenient to verify the reference values with D. Bazin.+The high-voltage of the [[Detectors#Tracking Parallel Plate Avalanche Counters (TPPAC)|TPPACs]], [[Detectors#Ionization Chamber|Ion Chamber]], the [[Detectors#Cathode Readout Drift Chambers (CRDC)|CRDCs]], [[Detectors#Plastic scintillators|FP Scintillator]], and [[Detectors#Hodoscope|Hodoscope]] are remotely controlled via GUI control panel (see figure below). 
 +The GUI can be opened from the desktop of the u4pc1 computer. Logging in as s800 userstart the singularity container for Debian 8 (Jessie) with
  
-{{:wiki:hv-gui.png?700|GUI control panel of the S800 detectors HV}}+  $ jessie_load
  
 +and use the script
  
-===== HV Manual control ==== +  $ /usr/opt/iseg-beta/start.sh
-Apart from the detectors listed above, the voltages of the [[Detectors#Plastic Scintillators|object scintillator]]  +
- and the **electrostatic gate** of the [[Detectors#Cathode Readout Drift Chambers (CRDC)|CRDCs]] are adjusted manually+
  
-The bias for the object scintillator is controlled via the H.V. power supply {{:wiki:Manual_Canberra_HV_3002D.pdf|Canberra model 3002 D}} located in the crate to the right of the data-U6 (see figure below). The voltage will be adjusted at the beginning of the experiment. This scintillator has to be turned off or put on "standby" when entering the vault due to its sensitivity to light.+The GUI control panel looks like this when the detectors are biased:
  
-In order for the CRDCs to work, it is necessary to bias their electrostatic gateThis is done via the H.V. Tennelec power supply located on the left side of the object-scintillator Canberra power supply (see figure below). +{{:wiki:iseg_slot0.png|}}
- +
  
-{{:wiki:nim-crate2-du6.jpg?400|NIM crate 2 in data-U6}}+The channel list is found elsewhere. 
 +The actual HV modules are located on the top level of the S3 vault, under the FP chamber. 
 + 
 +{{:wiki:img_1369.png|}} 
 + 
 +Other examples of HV values used with different fragment beams are: 
 + 
 +* **CRDC1 and CRDC2**:  
 +   * For He-3 @ ~130 MeV/u: CRDC1 (Anode=1120 V, Drift=1000 V); CRDC2 (Anode=1120 V; Drift=1000 V) 
 +   * For Be-12 @ ~30 MeV/u: CRDC1 (Anode=970 V, Drift=1000 V); CRDC2 (Anode=970 V; Drift=1000 V) 
 +   * For Ar-34 @ ~80 MeV/u: CRDC1 (Anode=860 V, Drift=1000 V); CRDC2 (Anode=840 V; Drift=1000 V) 
 +   * For Zn-58 @ ~44 MeV/u: CRDC1 (Anode=740 V, Drift=1000 V); CRDC2 (Anode=740 V; Drift=1000 V) 
 +   * For Rb-74 @ ~40 MeV/u: CRDC1 (Anode=650 V, Drift=1000 V); CRDC2 (Anode=650 V; Drift=1000 V) 
 +   * For U-238 (88+) @ ~70 MeV/u: CRDC1 (Anode=570 V, Drift=1000 V); CRDC2 (Anode=570 V; Drift=1000 V)  
 +     
 +* **FP scintillator (E1 up and E1 down)**:  
 +   * For He-3 @ ~130 MeV/u: UP (1790 V); DOWN (1760 V) 
 +   * For Be-12 @ ~30 MeV/u: UP (1770 V); DOWN (1700 V) 
 +   * For Ar-34 @ ~80 MeV/u: UP (1580 V); DOWN (1510 V) 
 +   * For Zn-58 @ ~44 MeV/u: UP (1400 V); DOWN (1340 V)  
 +   * For U-238 (88+) @ ~70 MeV/u: UP (1180 V); DOWN (1280 V) 
 + 
 +* **IC**:  
 +   * Typically 200 V
  
hv_bias.txt · Last modified: 2023/10/17 09:40 by noji