 +Other examples of HV values used with different fragment beams are:
 +
 +* **CRDC1 and CRDC2**: 
 +   * For He-3 @ ~130 MeV/u: CRDC1 (Anode=1120 V, Drift=1000 V); CRDC2 (Anode=1120 V; Drift=1000 V)
 +   * For Be-12 @ ~30 MeV/u: CRDC1 (Anode=970 V, Drift=1000 V); CRDC2 (Anode=970 V; Drift=1000 V)
 +   * For Ar-34 @ ~80 MeV/u: CRDC1 (Anode=860 V, Drift=1000 V); CRDC2 (Anode=840 V; Drift=1000 V)
 +   * For Zn-58 @ ~44 MeV/u: CRDC1 (Anode=740 V, Drift=1000 V); CRDC2 (Anode=740 V; Drift=1000 V)
 +   * For Rb-74 @ ~40 MeV/u: CRDC1 (Anode=650 V, Drift=1000 V); CRDC2 (Anode=650 V; Drift=1000 V)
 +   * For U-238 (88+) @ ~70 MeV/u: CRDC1 (Anode=570 V, Drift=1000 V); CRDC2 (Anode=570 V; Drift=1000 V) 
 +    
 +* **FP scintillator (E1 up and E1 down)**: 
 +   * For He-3 @ ~130 MeV/u: UP (1790 V); DOWN (1760 V)
 +   * For Be-12 @ ~30 MeV/u: UP (1770 V); DOWN (1700 V)
 +   * For Ar-34 @ ~80 MeV/u: UP (1580 V); DOWN (1510 V)
 +   * For Zn-58 @ ~44 MeV/u: UP (1400 V); DOWN (1340 V) 
 +   * For U-238 (88+) @ ~70 MeV/u: UP (1180 V); DOWN (1280 V)
 +
 +* **IC**: 
 +   * Typically 200 V
  