Line 50: | Line 50: | ||
* After the rough valve **I260RV** is automatically open, the mechanical pump will start pumping. | * After the rough valve **I260RV** is automatically open, the mechanical pump will start pumping. | ||
* The pressures in the mechanical pump (from **I259PG**) and beam chamber (from **I261PG**) will jump momentarily to very high values. They should go down slowly after few seconds (otherwise, the chamber is not properly closed). | * The pressures in the mechanical pump (from **I259PG**) and beam chamber (from **I261PG**) will jump momentarily to very high values. They should go down slowly after few seconds (otherwise, the chamber is not properly closed). | ||
- | * When pressure in the chamber (from **I261PG**) gets below 100 mTorr, **I260RV** will close. Foreline valve **I260FV** will open followed by the turbo gate valve **I260GV**. | + | * When pressure in the chamber (from **I261PG**) gets below 100 mTorr (at least 30 min), **I260RV** will close. Foreline valve **I260FV** will open followed by the turbo gate valve **I260GV**. |
- | * The pressures in the mechanical pump (from **I259PG**) and beam chamber (from **I261PG**) will jump momentarily to very high values. They should go down after few seconds | + | * Wait for few hours until vacuum is good. You can then turn on the ion gauge **I264IG** |
- | * Wait for some hours until vacuum is good | + | |