installation_fp_sci [2019/07/01 09:04]
pereira [Vent focal-plane chamber]
installation_fp_sci [2019/07/01 09:08]
pereira [Vent focal-plane chamber]
Line 12: Line 12:
 {{:wiki:NewPanelMate-S800FPVac.jpg?950|Panel Mate Vacuum FP chamber}} {{:wiki:NewPanelMate-S800FPVac.jpg?950|Panel Mate Vacuum FP chamber}}
   * Close I265GV (gate valve separating FP box from rest of beam line) is closed.   * Close I265GV (gate valve separating FP box from rest of beam line) is closed.
-  * Click on the **VENT STATUS** window and begin the venting procedure. The **VENT VALVE I260VV** window will turn red and display "OPEN". The pressure should remain constant because the manual venting valve is still closed. The turbo GV I260GV will be closed, isolating the turbo pump from the focal-plane chamber.+  * Click on the **VENT STATUS** window and begin the venting procedure. The **VENT VALVE I260VV** window will turn red and display "OPEN". The pressure should remain constant because the manual venting valve is still closed. The turbo GV **I260GV** will be closed, isolating the turbo pump from the focal-plane chamber.
   * Gently open the manual venting valve until the pressure displayed in the Pirani Gauge (see figure below) increases by ~1 torr every 15 seconds (see figures below)    * Gently open the manual venting valve until the pressure displayed in the Pirani Gauge (see figure below) increases by ~1 torr every 15 seconds (see figures below) 
 {{:wiki:fp-pirani.jpg?350|FP chamber manual venting valve}}  {{:wiki:fp-pirani.jpg?350|FP chamber manual venting valve}} 
   * Wait until the chamber is vented (few hours). The **I261PG** Pirani Gauge should read slightly above atmospheric pressure (due to the dry N2 flow pumped to the chamber). The window **VENT STATUS** in PanelMate should display "VENT VALVE OPEN". The window **PUMPDOWN STATUS** should display "PUMPDOWN STOPPED"   * Wait until the chamber is vented (few hours). The **I261PG** Pirani Gauge should read slightly above atmospheric pressure (due to the dry N2 flow pumped to the chamber). The window **VENT STATUS** in PanelMate should display "VENT VALVE OPEN". The window **PUMPDOWN STATUS** should display "PUMPDOWN STOPPED"
-  * Meanwhile, turn the turbo pump off using PanelMate and select **S800vac**+  * Meanwhile, turn the turbo pump off using PanelMate **S800vac**: select box **Turbo pump focal plane detector I260TP** can clock "Stop". It will take several minutes to fully stop (you can check the turbo-pump speed in box **I260TP SPEED**, or using EPICS channel **I260TP_FR_RD**) 
 +  * When the Turbo pump is fully stopped, switch off the chiller located near the S800 electronics rack
   * When the chamber is fully vented, select **VENT STATUS** and click **STOP VENTING PROCEDURE**. This will automatically close the vent valve **I260VV**   * When the chamber is fully vented, select **VENT STATUS** and click **STOP VENTING PROCEDURE**. This will automatically close the vent valve **I260VV**
   * Close the manual vent valve   * Close the manual vent valve
installation_fp_sci.txt · Last modified: 2023/08/18 10:19 by noji