 ===== Zeroing PTs ===== ===== Zeroing PTs =====
  
-  * When readings of the PTs (PT1 [S800_GHS:CMG_D2708A:VP_RD], PT2 [S800_GHS:CMG_D2703A:VP_RD], and PT3 [S800_GHS:CMG_D2708C:VP_RD]) are off (e.g. 1 Torr when pumped down), these need to be zeroed.+  * When readings of the capacitance manometers (PT1 [S800_GHS:CMG_D2708A:VP_RD], PT2 [S800_GHS:CMG_D2703A:VP_RD], and PT3 [S800_GHS:CMG_D2708C:VP_RD]) are off (e.g. 1 Torr when pumped down), these need to be zeroed.
   * The relevant PVs are "ZERO_CMD" instead of "VP_RD", e.g. S800_GHS:CMG_D2708A:ZERO_CMD for PT1. Send "1" to zero these PTs.   * The relevant PVs are "ZERO_CMD" instead of "VP_RD", e.g. S800_GHS:CMG_D2708A:ZERO_CMD for PT1. Send "1" to zero these PTs.
 +  * **Before “zeroing” a pressure gauge. Make sure it is completely bottomed out.** This would correspond to the GHS being in “Vent” mode and the TMP being on. If the IG is on and shows that the system is under High Vacuum (<10-3 Torr) you can go ahead and zero the gauge.
 +  * Here are instructions by Josh:
 +    * {{::zeroingpressuregauges1.png|}}
 +    * {{::zeroingpressuregauges2.png|}}
 +    * {{::zeroingpressuregauges3.png|}}
  