User Tools

Site Tools


nmr_probes

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision Both sides next revision
nmr_probes [2019/10/14 00:28]
pereira [Manually setting Dipole Fields]
nmr_probes [2019/10/17 10:29]
pereira [Manually setting Dipole Fields]
Line 85: Line 85:
   * Get access to the NMR controller in the mezzanine level of S3 vault (see figure below). Push the “SEARCH LOCAL” button to take control out of REMOTE mode.   * Get access to the NMR controller in the mezzanine level of S3 vault (see figure below). Push the “SEARCH LOCAL” button to take control out of REMOTE mode.
   * Select the right probe by turning the “probe knob” (see figure below). Remember, each dipole has two probes: probe 4, covers fields in the range 0.34 – 1.05 T, and probe 5 covers fields in the range 0.7–2.1 T. In this case, we should set the knob of the Spectrograph controller to position B    * Select the right probe by turning the “probe knob” (see figure below). Remember, each dipole has two probes: probe 4, covers fields in the range 0.34 – 1.05 T, and probe 5 covers fields in the range 0.7–2.1 T. In this case, we should set the knob of the Spectrograph controller to position B 
-        * The list of probes that can be selected with the knobs are, for the Spectrograph: A = D1 #4 probe; B= D1 #5; C=D2 #4; D=D2 #5. Analysis line: A = D1 #4 probe; B=D1 #5; C=D2 #4; D=D2 #5; E = D3 #4; F=D3 #5; G=D4 #4; H=D4 #5.+        * The list of probes that can be selected with the knobs are, for the Spectrograph: A = dipole D1probe 4; B= D1 probe #5; C=D2 #4; D=D2 #5. Analysis line: A = D1 #4 probe; B=D1 #5; C=D2 #4; D=D2 #5; E = D3 #4; F=D3 #5; G=D4 #4; H=D4 #5.
   * Use the “coarse” and “fine” knobs to select the field that you want (in T): 1.35454 T. By doing this, you are asking the NMR controller to just look at that field.    * Use the “coarse” and “fine” knobs to select the field that you want (in T): 1.35454 T. By doing this, you are asking the NMR controller to just look at that field. 
   * If the field of that dipole happens to be close enough to the field that the NMR controller is looking at, Bingo! You’ll most likely see a resonance in the oscilloscope. In the most likely situation, however, your magnet is far away from the NMR search window and you will see only a flat line in the oscilloscope.   * If the field of that dipole happens to be close enough to the field that the NMR controller is looking at, Bingo! You’ll most likely see a resonance in the oscilloscope. In the most likely situation, however, your magnet is far away from the NMR search window and you will see only a flat line in the oscilloscope.
nmr_probes.txt · Last modified: 2024/01/29 10:51 by swartzj