 ==== Manually setting Dipole Fields  ==== ==== Manually setting Dipole Fields  ====
 (Based on notes from Mathias Steiner with some extra info included) (Based on notes from Mathias Steiner with some extra info included)
 +
 +You will need to get access to the NMR controllers in the mezzanine level of S3 (see picture below). You will also have to connect an oscilloscope to one of the NMR controllers, either analysis line (left) or Spectrograph (right). Connect the left BNC connector (field modulation) to CH1 in the oscilloscope and the right BNC connector (NMR signal) in CH2. Select XY display, and AC coupling.
 +
 +
  
 Let’s assume that you want to set the dipole I265DS (first dipole (D1) of S800 Spectrograph) to the rigidity (Brho) 3.7965 Tm (this works for any dipole in the S800) Let’s assume that you want to set the dipole I265DS (first dipole (D1) of S800 Spectrograph) to the rigidity (Brho) 3.7965 Tm (this works for any dipole in the S800)