nmr_probes [2024/01/18 13:40]
swartzj [Error message: /srv/s3/nmr-logs/spectrograph.log is not accessible]
nmr_probes [2024/01/18 13:55]
swartzj [NMR reading sometimes takes too long (> 30 min)]
Line 124: Line 124:
   * This can be alleviated by changing the NMR thresholds (it can be done for each magnet in the analysis line, or for both magnets in the spectrograph).   * This can be alleviated by changing the NMR thresholds (it can be done for each magnet in the analysis line, or for both magnets in the spectrograph).
   * Stop NMR; enter new threshold (in Volts) and click “Set”; start NMR.   * Stop NMR; enter new threshold (in Volts) and click “Set”; start NMR.
-  * In general, a threshold 0.15 mV is good for normal NMR signals, whereas 0.10 mV may be necessary if the signal are too weak.+  * In general, a threshold 0.15 mV is good for normal NMR signals, whereas 0.10 mV may be necessary if the signals are too weak.
  
 2. Initial field where NMR starts scanning is far away from actual value (this has only been seen with Spectrograph) 2. Initial field where NMR starts scanning is far away from actual value (this has only been seen with Spectrograph)