==== Manual rebooting ==== | ==== Manual rebooting ==== | ||
In the rare case that the RunControl GUI gets " | In the rare case that the RunControl GUI gets " | ||
+ | |||
+ | ==== For hodoscope runs, cannot read signal from hodoscope ==== | ||
+ | If you see this kind error message when starting the DAQ with the hodoscope in the trigger (Ext. 1 in ULM gui), check the voltages on the crate which holds the hodoscope shaper modules. If they are not at the correct values, follow this procedure: | ||
+ | - Ordered List ItemTurn crate off | ||
+ | - Unplug both shapers (for both hodoscopes). Just pull them but don’t fully remove the cards. | ||
+ | - With shapers unplugged, turn crate on and check voltages | ||
+ | - With crate on, firmly push one of the shapers in. | ||
+ | - Push second shaper in | ||
+ | - Verify crate voltages | ||
+ | This procedure works for the electronic configuration with CAEN shaper modules. | ||
+ | |||
+ | |||