User Tools

Site Tools


to-do_list

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
to-do_list [2015/10/21 10:49]
pereira [In data U6]
to-do_list [2019/03/13 21:24] (current)
pereira [In data U6]
Line 22: Line 22:
   * Start [[Software#​u6pc5 (data U6)|Software applications]] (in computer u6pc5)   * Start [[Software#​u6pc5 (data U6)|Software applications]] (in computer u6pc5)
  
-  * Verify that **stagearea** in u6pc5 is pointing to the right place. After typing ''​ls -l ~/​stagearea''​ in a u6pc5 terminal, you should see: **/​user/​s800/​stagearea -> /​mnt/​daqtesting/​s800/​exxxxx**,​ where xxxxx corresponds ​to the experiment number. If the experiment number is not right, follow the following steps+  * Source ​the environment file to run the DAQ
-        * Create ​experiment directory ​in **/​mnt/​daqtesting/​s800** by typing ​''​mkdir exxxxx''​ +        * Open terminal ​in u6pc5 and type ''​ssh -Y spdaq50''​ 
-        * Go to the newly created experiment directory, and create subdirectories **s800built**,​ **rawccusb**,​ and **rawvmusb** +        * Type ''​/usr/opt/nscldaq/xxxx/​daqsetup.bash''​, where xxxx is the nscldaq version number (11.0-015 on Oct 262015)
-        * Create the stagearea link by typing ​''​ln -s /mnt/daqtesting/s800/exxxxx stagearea'' ​ +
-        * Run the S800 DAQ with multilogger enabledand take some data on disk to ensure that the newly created stagearea is properly filled+
  
 +  * Verify with S800 Device Physicist that S800 DAQ has been installed and it is ready for the experiment. S800 DAQ can be installed by typing command ''​goinstalldaq''​ and following the script instructions (please, contact the S800 Device Physicist if you are not sure of what you're doing).
  
-  * Check scalers while running S800 DAQ 
  
 +  * Verify that **stagearea** in u6pc5 is pointing to the right place. After typing ''​ls -l ~/​stagearea''​ in a u6pc5 terminal, you should see: **/​user/​s800/​stagearea -> /​mnt/​events/​operations/​s800/​exxxxx**,​ where xxxxx corresponds to the experiment number. If the experiment number is not right, contact S800 Device Physicist.
 +
 +  * Set S800 DAQ in [[s800 daq tools#​Running in Slave mode with multilogger |Slave mode and enable multilogger]] to record S800 data in the s800 stagearea.
 + 
 +  * Check [[s800 daq tools#​scaler display|scalers]] while running S800 DAQ
  
   * [[Turning magnets on/off#Set magnet currents individually |Put 10 Amps in Spectrograph dipoles]]   * [[Turning magnets on/off#Set magnet currents individually |Put 10 Amps in Spectrograph dipoles]]
Line 41: Line 44:
    
   * [[Magnetic-rigidity setting|Set spectrograph Brho]] (coordinate with experimenters)   * [[Magnetic-rigidity setting|Set spectrograph Brho]] (coordinate with experimenters)
 +
 +
 +
 +
 +===== At the end of the experiment =====
 +  * Verify with S800 Device Physicist that S800 DAQ software is uninstalled. This can be done by typing script ''​gouninstalldaq''​ (please, contact Device Physicist if you are not sure of what you're doing)
 +
 +  * Unbias detectors, CRDC interface boards, CRDCs gate, etc
 +
 +  * [[Gas handling system|Remove gas]] from CRDCs and Ion Chamber ​
  
  
  
to-do_list.1445438989.txt.gz · Last modified: 2015/10/21 10:49 by pereira