User Tools

Site Tools


tuning_the_s800_xdt

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision Both sides next revision
tuning_the_s800_xdt [2017/07/18 11:48]
pereira [CRDCs setup]
tuning_the_s800_xdt [2017/07/21 17:35]
pereira [CRDCs setup]
Line 135: Line 135:
   * Check **[[s800 SpecTcl|Spectcl]]** window **S800_CRDCS.win** (see figure below) to verify the good performance of the detectors. (The spectra for each CRDC can be checked separatelly in windows **s800_CRDC1.win** and **S800_CRDC2.win**)   * Check **[[s800 SpecTcl|Spectcl]]** window **S800_CRDCS.win** (see figure below) to verify the good performance of the detectors. (The spectra for each CRDC can be checked separatelly in windows **s800_CRDC1.win** and **S800_CRDC2.win**)
  
-      * Spectra **crdc1.raws** and **crdc2.raws** (top and middle spectra in the leftmost (first) column)+      * Spectra **crdc1.raws** and **crdc2.raws** 
           * Each spectra shows the multiple sampled signals from each pad            * Each spectra shows the multiple sampled signals from each pad 
           * These are good spectra to check if the pad thresholds are properly set. Thresholds are too low if you see that all pads are firing at low energies. If that's the case, increase the pad thresholds in files **s800crdcv1.tcl** and **s800crdcv2.tcl**, under directory **s800/operations/daq/usb/Configs** (contact device physicist for assistance, if needed).           * These are good spectra to check if the pad thresholds are properly set. Thresholds are too low if you see that all pads are firing at low energies. If that's the case, increase the pad thresholds in files **s800crdcv1.tcl** and **s800crdcv2.tcl**, under directory **s800/operations/daq/usb/Configs** (contact device physicist for assistance, if needed).
Line 143: Line 143:
           * Adjust anode HV to bring fuzzy maximum to around 600-700 channels (saturation of each pad at ~ 1000 ch)           * Adjust anode HV to bring fuzzy maximum to around 600-700 channels (saturation of each pad at ~ 1000 ch)
  
-      * Spectra **crdc1.anode_crdc1.tac** and **crdc2.anode_crdc2.tac** (top and middle spectra in the second column)+      * Spectra **crdc1.anode_crdc1.tac** and **crdc2.anode_crdc2.tac** 
           * This spectrum is used to ensure that the field in the detectors is uniform and well aligned. If the detector is working properly, then the amplitude of the ANODE signals should not depend on the position of the beam (as shown in the figure below)            * This spectrum is used to ensure that the field in the detectors is uniform and well aligned. If the detector is working properly, then the amplitude of the ANODE signals should not depend on the position of the beam (as shown in the figure below) 
           * If the field is not well aligned, then there will be a correlation between ANODE signals and TAC signals            * If the field is not well aligned, then there will be a correlation between ANODE signals and TAC signals 
  
-     * Spectra **crdc1.x_crdc1.padsum** and **crdc2.x_crdc2.padsum** (top and middle spectra in the third column)+     * Spectra **crdc1.x_crdc1.padsum** and **crdc2.x_crdc2.padsum** 
           * It shows the sum over multiple sampled signals from each pad along the (dispersive) x position            * It shows the sum over multiple sampled signals from each pad along the (dispersive) x position 
           * The padsum signals should not show any correlation with the x (dispersive) position                 * The padsum signals should not show any correlation with the x (dispersive) position      
  
-      * Spectra **crdc1.xg_crdc1.tac** and **crdc2.xg_crdc2.tac** (top and middle spectra in the fourth column)+      * Spectra **crdc1.xg_crdc1.tac** and **crdc2.xg_crdc2.tac** 
           * It shows the beam distribution in the dispersive (xg) //vs// non-dispersive (tac) directions           * It shows the beam distribution in the dispersive (xg) //vs// non-dispersive (tac) directions
           * It is used to ensure that the fragment beam is centered in the detectors           * It is used to ensure that the fragment beam is centered in the detectors
           * It is also used to see the effect of the beam blocker (used to stop intense contaminants) in the cocktail beam                  * It is also used to see the effect of the beam blocker (used to stop intense contaminants) in the cocktail beam       
      
-      * Spectra **crdc1.xg** and **crdc2.xg** (bottom spectra in first and second columns)+      * Spectra **crdc1.xg** and **crdc2.xg** 
           * It shows the position of the beam in the dispersive direction, evaluated by calculating the "center of gravity". The peak should be in the middle of the spectra in order to center the beam           * It shows the position of the beam in the dispersive direction, evaluated by calculating the "center of gravity". The peak should be in the middle of the spectra in order to center the beam
  
- +      * Spectra **crdc1.tac** and **crdc2.tac** 
-      * Spectra **crdc1.tac** and **crdc2.tac** (bottom spectra in third and fourth columns)+
           * They correspond to the non-dispersive position of the beam in the CRDCs.            * They correspond to the non-dispersive position of the beam in the CRDCs. 
 +
 +      * Spectra **crdc1.pad_mult** and **crdc2.pad_mult** 
 +          * They show the distribution of pad multiplicies (i.e. number of pads firing for each event). Typical average values are ~10-15. Significantly larger average multiplicities might indicate that the thresholds are set too low.
  
 {{:wiki:CRDCS-example.png?850|S800_CRDCS.win SpecTcl window}} {{:wiki:CRDCS-example.png?850|S800_CRDCS.win SpecTcl window}}
tuning_the_s800_xdt.txt · Last modified: 2023/09/22 15:15 by swartzj