Line 46: Line 46:
  
 **Important note**: Ensure that the experimenters never pump the chamber down if the spectrograph vacuum section is vented. The pressure gradient between both sections would "push" the structure of the spectrograph beam line towards the target chamber, potentially damaging the bellows of the line. **Important note**: Ensure that the experimenters never pump the chamber down if the spectrograph vacuum section is vented. The pressure gradient between both sections would "push" the structure of the spectrograph beam line towards the target chamber, potentially damaging the bellows of the line.
 +
 +
 ===== Spectrograph ===== ===== Spectrograph =====
 This system is pumped with cryopumps. Please do not vent this section if the target chamber is under vacuum pressure. The pressure gradient between both sections would "push" the structure of the spectrograph beam line towards the target chamber, potentially damaging the bellows of the line. This system is pumped with cryopumps. Please do not vent this section if the target chamber is under vacuum pressure. The pressure gradient between both sections would "push" the structure of the spectrograph beam line towards the target chamber, potentially damaging the bellows of the line.
vacuum_control.txt ยท Last modified: 2019/10/17 10:43 by pereira