Line 53: | Line 53: | ||
- Using the EMIS controllers in the middle level of the S3 vault (see figure above), ensure that the ion gauge I225IG is turned off. | - Using the EMIS controllers in the middle level of the S3 vault (see figure above), ensure that the ion gauge I225IG is turned off. | ||
- Ensure that gate valves I246GV (between lower-dipole section and target area) and I215GV (between intermediate image area and the lower-dipole section) are closed. | - Ensure that gate valves I246GV (between lower-dipole section and target area) and I215GV (between intermediate image area and the lower-dipole section) are closed. | ||
- | - Open S800vac | + | - Open **S800vac** PanelMate. |
- Go to page 4 (" | - Go to page 4 (" | ||
- Window "VENT STATUS" | - Window "VENT STATUS" |