venting_pumping_intermediate_image_box

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Next revision Both sides next revision
venting_pumping_intermediate_image_box [2018/12/19 11:43]
pereira [Filling/Removing gas in Intermediate-image delay PPAC]
venting_pumping_intermediate_image_box [2023/02/13 10:44]
noji Updated descriptions about the II-GHS based upon "S800 Intermediate Image PPAC Gas Handling System Operation Manual" (v4) provided to us on 2/8/2023 by B. Ewert.
Line 6: Line 6:
  
 ===== General information ===== ===== General information =====
-The procedures to vent/pump the S800 Intermediate Image box, and/or filling the PPAC detectors with gas must be done carefully to ensure that the PPACs are not damaged. For instance, before venting the chamber, we have to make sure that the PPACs are empty and equalized with respect to the vacuum in the chamber. In order to avoid errors, an automated gas handling system (II-GHS) was designed to perform these tasks safely. The system is schematically illustrated in the figure below. It shows the status of the valves (red: closed, white: open) when the intermediate image chamber is under vacuum and the PPACs are filled with gas (NOTE: This figure was taken from the A1900 GHS included in Barney. A similar GUI for the S800 II PPACs is not yet available): +The procedures to vent/pump the S800 Intermediate Image box, and/or filling the PPAC detectors with gas must be done carefully to ensure that the PPACs are not damaged. For instance, before venting the chamber, we have to make sure that the PPACs are empty and equalized with respect to the vacuum in the chamber. In order to avoid errors, an automated gas handling system (II-GHS) was designed to perform these tasks safely. The II-GHS equipment (valves, flow-meter, pipes, isobutane gas bottle. etc.) are located on the west balcony of the top level in the S3 vault (see figure below):
- +
-{{:wiki:ghsiippacs.png?1000|Gas Handling System of S800 Intermediate Image}}  +
- +
-The II-GHS equipment (valves, flow-meter, pipes, isobutane gas bottle. etc.) are located on the west balcony of the top level in the S3 vault (see figure below):+
  
 {{:wiki:IIGHShardware.jpg?600|Gas Handling System of S800 Intermediate Image}}  {{:wiki:IIGHShardware.jpg?600|Gas Handling System of S800 Intermediate Image}} 
  
 +The GHS is operated remotely through CS studio. Its page can be found on the EXP network through FRIB EXP Main → S800 → GHS Pages (Secondary Pages) → Intermediate Image PPAC and is shown below. 
  
 +{{:wiki:ii-ppac-ghs-css.png|}}
  
 +The layout of the II-GHS is shown below.
  
 +{{:wiki:ii-ppac-ghs-layout.png|}}
  
 +The operational states of the II-GHS are shown in the diagram below.
  
 +{{:wiki:ii-ppac-ghs-operationalstates.png|}}
  
-===== Vent Intermediate-image chamber ===== +===== Emptying Detectors ===== 
-Requirements: PanelMate files **S3_Gas_Handling** and **S800Vac** +  - Set “PPAC Setpoint” to 0 Torr. 
- +  - Click **EMPTY** on the “GHS_II-PPAC_Main” page to begin pumping out the gas quickly
-  * Verify that PPACs are in automatic mode with gas removed (**S3_Gas_Handling**) +  When “PPAC Pressure” reads < 0.1 Torr, click **MAN** and then **AUTO**. This will release the remaining gas into the diagnostics box. 
-         The "GH System Status box should read "Auto Mode Gas Removed" +  - Monitor D2340 pumping station and ensure the TMP stays on and TGV remains openIf not, contact the Vacuum Operation group
-        If gas is removed but the system is not in the automatic mode, consult the Beam Physicists or Device Physicist. NOTE: the only valves that should be open in **S3_Gas_Handling** are the "Det Bypass" and "Chmbr Bypass  +  - Valves #1-5 should be closed and valve #6 should remain openFrom this state click PUMP to enter pump mode and leave the system in this state until it is ready to be filled. 
-  * Verify that PPACs are not biased: +    Jorge noted on 2/12/2023 that this step did not work; he reported that when he clicked **PUMP** nothing happened. Insteadhe had to “manually” enter the PV name associated with this command and changed from 0 to 1He speculated that this was most likely due to a CS Studio problem.
-        * Check that the detectors are unbiased by opening the [[HV bias|HV GUI]] +
-        * Notify the system of the bias-off state by selecting the "I214BON_CONFIRM" box in **S3_Gas_Handling** and clicking "PP BIAS OFF CONFIRMED" +
-  * Open **S800Vac** and go to page 03 ("Intermediate image box vacuum system sec 15") +
-  * Isolate chamber by closing gate valves **I212GV** and **I215GV** +
-  * [[Vacuum control|Turn off]] ion gauge **I213IG**  +
-  * Check box "PUMPDOWN STATUS INTERMEDIATE IMAGE BOX". It should read "Good vacuum"  +
-{{:wiki:iis800vac_pumped.jpg?500|S800vac Panel Mate}}.  +
-  * In order to begin venting process, select box "VENT STATUS" and click "BEGIN VENTING PROCEDURE" +
-        * Box "VENT STATUS" should change from "VENT PROCEDURE STOPPED" to "WAITING FOR VENT TIMER".  +
-        * Turbo gate valve **I214GV** will be automatically closed and pressure in chamber will start rising (as seen in **I215PG**) {{:wiki:iis800vac_startventing.jpg?500|S800vac Panel Mate: start venting}}.  +
-        * When **I215PG** reads atmospheric pressureselect the box "VENT STATUS" and click "STOP VENTING PROCESS". {{:wiki:iis800vac_stopventing.jpg?500|S800vac Panel Mate: stop venting}}.  +
- +
-        The status in the box "VENT STATUS" should change from "WAITING FOR VENT TIMER" to "VENT PROCEDURE STOPPED" +
-        You can now open the chamber +
- +
- +
-===== Pumping Intermediate-image chamber ===== +
-Requirements: PanelMate files **S3_Gas_Handling** and **S800Vac** +
-  * Verify that Chamber is closed and pressure-relief valve (north side of the chamber) is properly closed (NOTE: sometimes, due to overpressure from dry N2 during the venting procedure, the O-ring in pressure-relief valve flange is moved out of place) +
-  * Verify that PPACs are not biased by opening the [[HV bias|HV GUI]] +
-  * IMPORTANT: Verify that manual roughing pump valve located near the intermediate-image chamber is closed{{:wiki:manualvalverough.jpg?350|Manual valve rough pump}}.  +
- +
-  * Insert PPACs/turn on light (Diagnostic Drives PanelMate) and verify condition through sight window on south side of the chamber (be careful no to bump the bullet camera). +
-  * Open **S800Vac** and go to page 03 ("Intermediate image box vacuum system sec 15") +
-  * Check "PUMPDOWN STATUS INTERMEDIATE IMAGE BOX"; it should read "PUMPDOWN STOPPED"Select that box and click on "START PUMPDOWN". The chamber will not be pumped yet because the manual roughing pump valve is closed+
-  * Slowly crack manual roughing valve +
-        * Start slow (< Torr/sec). Chamber pressure is displayed in **I215PG** in **S800Vac** PanelMate. There is also a local display of the mechanical pump pressure on a Granville-Phillips 275 Mini-Convectron, which is helpful to control rate of pumping. +
-        * Verify PPACs changing shape through sighting window. +
-        * Pump to about 100 Torr before fully opening the roughing valve. +
-  * Once roughing valve is fully open, retract PPACs and turn lights off. +
-  * When the pressure in **I215PG** is below 200 mTorr, the Turbo Pump will start automatically through the following sequence: +
-        * **I214RV** will be closed +
-        **I214FV** will be opened +
-        * **I214GV** will be opened +
-  * When the pressure ready by **I215PG** is few mTorr, you can turn on the ion gauge **I213IG**. Good vacuum will take ~1-hours. +
-  * NOTE: it sometimes happens that the foreline pressure rises above the trip point for the Turbo Pump (e.g. if the pumpdown timer expires before pressure is below 300 mTorr, or there is too much back-up pressure from the chamber when the foreline valve opens). In this situation, you will see that: (1) the turbo pump will stop, (2) the turbo gate valve **I214GV** will be closed(3) the foreline valve **I214FV** will be closed and the corresponding PanelMate box will turn red and display an error message "Foreline pressure > 2 Torr", and (4) the foreline pressure will be few Torr. In a situation like this, foreline valve needs to be bypassed open (in PanelMate) and the TP needs to be restarted: +
-          * Select box labeled "FORELINE VLVE I214FV" and the click on "BYBASS VAC INTERLOCKS OPEN" button +
-          * Select box labeled "TURBO PUMP INTERMEDIATE IMAGE BOX I214TP" and  click "START" +
-          * Turbo pump will start ramping speedWhen the TP reaches 40000 RPM, the **I214GV** will open automatically. +
-          * When foreline vacuum gets below 2 Torr, you can select "FORELINE VLVE I214FV" and click "REMOVE BYPASS" +
-          * (Note: sometimes it is necessary to repeat this operation couple of times).  +
-  * At the end of the procedure, close the manual roughing pump valve  +
- +
- +
- +
- +
-===== Filling/Removing gas in Intermediate-image delay PPAC ===== +
- +
-{{:wiki:ghsiippacs.png?1000|Gas Handling System of S800 Intermediate Image}} +
  
-Requirements: PanelMate files **S3_Gas_Handling** and **S800Vac**+===== Filling and Running Detectors ===== 
 +  - If in Vent Mode, click **PUMP** then click **RUN** to enter Run Mode. 
 +  - If in Pump Mode, click **RUN** to enter Run Mode. 
 +  - Plot and monitor IG_D2430:VP_RD while filling. Ion gauge reading will increase but should remain <1E-5 Torr when PPACs are running at full pressure.  
 +  - Step up “PPAC Setpoint” by 1 Torr from 0 to 6 Torr. Allow the “PPAC Pressure” to stabilize for 15 seconds at each step. This should be a quick process but **DO NOT** simply set the “PPAC Setpoint” to 6 Torr as the gas flow could overwhelm the system. 
 +  - PPAC pressure should stabilize after ~1 min. HV may now be turned on to operate the PPACs.
  
-==== Filling Intermediate-image delay PPAC with gas ==== +===== Troubleshooting ===== 
-  * Verify that PPACs are not biased: +  - If any issues are encountered during fillingrunningor emptyingcontact the Vacuum Operation group for assistance.
-        * Check that the detectors are unbiased by opening the [[HV bias|HV GUI]] +
-        * Notify the system of the bias-off state by selecting the "I214BON_CONFIRM" box in **S3_Gas_Handling** and clicking "PP BIAS OFF CONFIRMED" +
-  * Verify that PPACs are in automatic mode with gas removed (**S3_Gas_Handling**) +
-        * The "GH System Status box should read "Auto Mode Gas Removed" +
-        * If gas is removed but the system is not in the automatic mode, consult the Beam Physicists or Device Physicist. This typically requires to verify that the system is in a controlled status. In particular, the "I214BON_CONFIRM" box should indicate "PP BIAS OFF CONFIRMED", and the only that should be open in **S3_Gas_Handling** are the "Det Bypass" (V5) and "Chmbr Bypass" (V6); the rest of the valves (Gas Sup V1Gas Lin Pmp Out V2Flow Bypass V3Det Return V4, and I214V_NCL (enable) should be closed)  +
-  * Confirm that PanelMate box "I214P_AUTO/MAN Pre Set" reads 5 Torr (the set gas pressure in the detector)  +
-  * Open the gas bottle valve (sitting near the gas handling system) +
-  * PanelMate box "Gas Flow Control" should indicate "OK to Start Gas Flow"Select that box and click on "Start Gas Flow" +
-        * Valve V2 (Gas lin pmp out) will be opened and I214V_NCL will change from closed to "ENABLED" +
-        * Valves V3 (Flow Bypass) and V4 (Det Return) will open +
-        * Valve V1 (Gas supply) will open, and V2, V3, and V6 (Chamber Bypass) will close +
-        * GH System status should indicate "Gas Flowing" +
-        * I214CM Press Rdg will gradually increase to the set value +
-  * When the detector is filled, valve V5 (Det Bypass) will close +
-        * GH System status will indicate "Flow Estab." +
-        * I214BON_CONFIM will change to "PPAC BIAS ASSUMED ON" +
-        * The valves status will be: V1 (open), V2 (closed), V3 (closed), V4 (open), V5 (closed), V6 (closed, I214V_NCL (enabled) +
-        * You can now turn on the detectors+