This shows you the differences between two versions of the page.
Both sides previous revision Previous revision Next revision | Previous revision Next revision Both sides next revision | ||
venting_pumping_intermediate_image_box [2018/12/20 15:12] pereira [Filling/Removing gas in Intermediate-image delay PPAC] |
venting_pumping_intermediate_image_box [2019/02/13 09:40] pereira [Removing gas from Intermediate-image delay PPAC] |
||
---|---|---|---|
Line 98: | Line 98: | ||
* GH System status will indicate "Flow Estab." | * GH System status will indicate "Flow Estab." | ||
* I214BON_CONFIM will change to "PPAC BIAS ASSUMED ON" | * I214BON_CONFIM will change to "PPAC BIAS ASSUMED ON" | ||
- | * The valves status will be: V1 (open), V2 (closed), V3 (closed), V4 (open), V5 (closed), V6 (closed, I214V_NCL (enabled) | + | * The valves status will be: V1 (open), V2 (closed), V3 (closed), V4 (open), V5 (closed), V6 (closed), I214V_NCL (enabled) |
* You can now turn on the detectors | * You can now turn on the detectors | ||
Line 117: | Line 117: | ||
* Det Bypass V5 (closed) | * Det Bypass V5 (closed) | ||
* Chmbr Bypass (closed) | * Chmbr Bypass (closed) | ||
- | * I1214V_NCL | + | * I214V_NCL |
* PanelMate box "Gas Removal Control" | * PanelMate box "Gas Removal Control" | ||
* Valves V1 (I214 Gas Sup) and I214V_NCL will close | * Valves V1 (I214 Gas Sup) and I214V_NCL will close | ||
Line 129: | Line 129: | ||
* I214BON_CONFIM will change to "PPAC BIAS ASSUMED ON" | * I214BON_CONFIM will change to "PPAC BIAS ASSUMED ON" | ||
* The valves status will be: V1 (closed), V2 (closed), V3 (closed), V4 (closed), V5 (open), V6 (open), I214V_NCL (closed) | * The valves status will be: V1 (closed), V2 (closed), V3 (closed), V4 (closed), V5 (open), V6 (open), I214V_NCL (closed) | ||
- | + | * Close the gas bottle valve (sitting near the gas handling system) | |