Line 13: | Line 13: | ||
- Notify the Control Room that EXP alarms may occur. | - Notify the Control Room that EXP alarms may occur. | ||
- Ensure that beamline gate valves D2436 BGV (between intermediate image area and the lower-dipole section) and D2542 BGV (between lower-dipole section and target area) are closed. | - Ensure that beamline gate valves D2436 BGV (between intermediate image area and the lower-dipole section) and D2542 BGV (between lower-dipole section and target area) are closed. | ||
- | - Verify that the target area is vented. | ||
- Ensure that the ion gauge RBT_BTS33: | - Ensure that the ion gauge RBT_BTS33: | ||
- Check the box next to “Show Valve Controls”, | - Check the box next to “Show Valve Controls”, |