 +==== Pumping analysis-line lower-dipole section ====
 +  - Make sure that the manual gate valve is closed (see above).
 +  - Open **S800vac** PanelMate.
 +  - Go to page 4 ("lower beamline dipoles vacuum system, section 16")
 +  - Select box"PUMPDOWN STATUS SECTION 16-LOWER DIPOLES" and click "START PUMPDOWN". Good vacuum will take few hours.
 +  - When *I226PG* reads pressures of few mTorr, you can turn on the ion gauge **I225IG**.
 +
 +