This shows you the differences between two versions of the page.
Both sides previous revision Previous revision Next revision | Previous revision Next revision Both sides next revision | ||
venting_pumping_lower-dipole_section_venting_pumping_lower-dipole_section [2017/06/22 12:03] pereira [Pumping analysis-line lower-dipole section] |
venting_pumping_lower-dipole_section_venting_pumping_lower-dipole_section [2020/04/01 21:16] pereira [Venting analysis-line lower-dipole section] |
||
---|---|---|---|
Line 4: | Line 4: | ||
In order to ensure that the beamline in the target area is properly aligned with respect to the beam axis, it is necessary to install a telescope in between the last two quadrupoles of the analysis line, as well as a "cross hair" plate in between the last quadrupole and the target chamber. Although the alignment of the telescope is done by lab staff (e.g. Dave Sanderson), the user is responsible for removing the gate valve I246GV separating the analysis line and the target area. This of course, requires that the lower-dipole section is properly vented. The venting process is generally done by the device physicist, following the steps: | In order to ensure that the beamline in the target area is properly aligned with respect to the beam axis, it is necessary to install a telescope in between the last two quadrupoles of the analysis line, as well as a "cross hair" plate in between the last quadrupole and the target chamber. Although the alignment of the telescope is done by lab staff (e.g. Dave Sanderson), the user is responsible for removing the gate valve I246GV separating the analysis line and the target area. This of course, requires that the lower-dipole section is properly vented. The venting process is generally done by the device physicist, following the steps: | ||
- | | + | |
- | | + | |
- | | + | |
- | | + | * Ensure that gate valves I246GV (between lower-dipole section and target area) and I215GV (between intermediate image area and the lower-dipole section) are closed. |
- | - Go to page 4 (" | + | * Window "VENT STATUS" |
- | - Window "VENT STATUS" | + | |
- | | + | * In PanelMate **S800vac** (page 4), click on "VENT STATUS" |
- | | + | |
- | | + | |
- | | + | |