Line 13: | Line 13: | ||
- Notify the Control Room that EXP alarms may occur. | - Notify the Control Room that EXP alarms may occur. | ||
- Ensure that beamline gate valves D2436 BGV (between intermediate image area and the lower-dipole section) and D2542 BGV (between lower-dipole section and target area) are closed. | - Ensure that beamline gate valves D2436 BGV (between intermediate image area and the lower-dipole section) and D2542 BGV (between lower-dipole section and target area) are closed. | ||
- | - Verify that the target area is vented. | ||
- Ensure that the ion gauge RBT_BTS33: | - Ensure that the ion gauge RBT_BTS33: | ||
- Check the box next to “Show Valve Controls”, | - Check the box next to “Show Valve Controls”, | ||
Line 20: | Line 19: | ||
- After the section is vented (reached at atmospheric pressure), the vent valve should close automatically. | - After the section is vented (reached at atmospheric pressure), the vent valve should close automatically. | ||
- | ===== Pumping analysis-line lower-dipole section ===== | + | ===== Pumping-down analysis-line lower-dipole section ===== |
- The beamline gate valves D2436 BGV and D2542 BGV, the vent valve, the turbo gate valve, and the roughing valve (RV) should be all closed. The foreline valve (FV) should remain open. The ion gauge should remain turned off. The vacuum status should be red with "BAD VACUUM" | - The beamline gate valves D2436 BGV and D2542 BGV, the vent valve, the turbo gate valve, and the roughing valve (RV) should be all closed. The foreline valve (FV) should remain open. The ion gauge should remain turned off. The vacuum status should be red with "BAD VACUUM" |