 ====== Venting/pumping Object box ====== ====== Venting/pumping Object box ======
  
-===== Vent Object chamber ===== +In CS-Studio/Phoebus, from the S800 Main Vacuum page, go to the Object chamber page, opening the page shown below.
- +
-From the S800 Main Vacuum page, go to the Object chamber page, opening the page shown in the figure below.+
  
 {{:wiki:screenshot_from_2024-02-06_09-20-58.png|}} {{:wiki:screenshot_from_2024-02-06_09-20-58.png|}}
 +
 +===== Vent Object chamber =====
  
 The vent valve and the roughing valve RV should be closed when you begin.  The vent valve and the roughing valve RV should be closed when you begin. 
Line 11: Line 11:
 The venting procedure is as follows:  The venting procedure is as follows: 
   - Notify the Control Room and inform them that they can disregard the EXP alarm that will occur   - Notify the Control Room and inform them that they can disregard the EXP alarm that will occur
 +  - Make sure that the detectors inside the chamber have been unbiased
   - Close the beamline gate valves D2341 BGV and D2303 BGV   - Close the beamline gate valves D2341 BGV and D2303 BGV
   - Turn off the IG   - Turn off the IG