User Tools

Site Tools


venting_pumping_object_box

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
venting_pumping_object_box [2018/04/25 17:49]
pereira [Vent Object chamber]
venting_pumping_object_box [2020/04/01 21:32]
pereira [Vent Object chamber]
Line 11: Line 11:
   * [[Vacuum control|Turn off]] ion gauge **I188IG** ​   * [[Vacuum control|Turn off]] ion gauge **I188IG** ​
   * Check box "​PUMPDOWN STATUS BEAMLINE OBJECT BOX". It should read "Good vacuum" ​   * Check box "​PUMPDOWN STATUS BEAMLINE OBJECT BOX". It should read "Good vacuum" ​
 +  * Go to the S800 OBJ station and make sure that the manual valve (see figure below) is closed.
 +
 {{:​wiki:​objs800vac_pumped.jpg?​500|S800vac Panel Mate}}. ​ {{:​wiki:​objs800vac_pumped.jpg?​500|S800vac Panel Mate}}. ​
   * In order to begin venting process, select box "VENT STATUS"​ and click "BEGIN VENTING PROCEDURE"​   * In order to begin venting process, select box "VENT STATUS"​ and click "BEGIN VENTING PROCEDURE"​
Line 16: Line 18:
         * Turbo gate valve **I188GV** will be automatically closed         * Turbo gate valve **I188GV** will be automatically closed
         * Vent valve **I190VV** will be automatically opened. However, the pressure in the chamber (ass seen in **I191PG** will not change because we need to open the manual venting valve         * Vent valve **I190VV** will be automatically opened. However, the pressure in the chamber (ass seen in **I191PG** will not change because we need to open the manual venting valve
-        * Go to the S800 OBJ station and slowly open the manual valve (see figure below). You can see the pressure rising in the Granville-Phillips 275 Mini-Convectron display.+        * Go to the S800 OBJ station and slowly open the manual valve (see figure below). You can see the pressure rising in the Granville-Phillips 275 Mini-Convectron display ​located on top of the OBJ box.
         * When **I191PG** reads atmospheric pressure select the box "VENT STATUS"​ and click "STOP VENTING PROCESS"​         * When **I191PG** reads atmospheric pressure select the box "VENT STATUS"​ and click "STOP VENTING PROCESS"​
         * The status in the box "VENT STATUS"​ should display "VENT PROCEDURE STOPPED"​ and the vent valve **I190VV** should be automatically closed         * The status in the box "VENT STATUS"​ should display "VENT PROCEDURE STOPPED"​ and the vent valve **I190VV** should be automatically closed
venting_pumping_object_box.txt ยท Last modified: 2020/04/01 21:32 by pereira