This shows you the differences between two versions of the page.
Both sides previous revision Previous revision Next revision | Previous revision | ||
venting_pumping_object_box [2018/05/03 09:10] pereira |
venting_pumping_object_box [2023/06/02 17:14] (current) swartzj |
||
---|---|---|---|
Line 2: | Line 2: | ||
===== Vent Object chamber ===== | ===== Vent Object chamber ===== | ||
- | Requirements: | ||
- | * Open **S800Vac** and go to page 01 (" | + | From the S800 Main Vacuum page, go to the Object chamber |
- | * Verify that OBJ SCI is not biased: | + | * Before you start, make sure that the gate valves |
- | * Check that the detectors is unbiased | + | * The vent valve should |
- | * Notify the system | + | * The valve to the turbo pump TVG should be closed. |
- | * Isolate chamber by closing | + | |
- | * [[Vacuum control|Turn | + | |
- | * Check box " | + | |
- | {{: | + | |
- | * In order to begin venting process, select box "VENT STATUS" | + | |
- | * Box "VENT STATUS" | + | |
- | * Turbo gate valve **I188GV** will be automatically | + | |
- | * Vent valve **I190VV** will be automatically opened. However, the pressure in the chamber | + | |
- | * Go to the S800 OBJ station and slowly open the manual valve (see figure below). You can see the pressure rising in the Granville-Phillips 275 Mini-Convectron display located on top of the OBJ box. | + | |
- | * When **I191PG** reads atmospheric pressure select the box "VENT STATUS" | + | |
- | * The status in the box "VENT STATUS" | + | |
- | * Close the manual valve | + | |
- | * You can now open the chamber | + | |
- | {{:wiki: | + | The procedure is then as follows: |
+ | - Close the valve to the roughing pump RV. | ||
+ | - Now you can plot the pressure vs time at BPG to monitor the pressure before you open the vent valve. | ||
+ | - **Keep a constant eye on the pressure level** as it increases rapidly. When it approaches 800 torr, **it is imperative that you close the venting valve.** Failure to do so will result in over-pressure in the chamber since it will continue to be filled with dry nitrogen. | ||
+ | - You can now enter and open the chamber. | ||
+ | ===== Pumping Object chamber ===== | ||
- | + | From S800 Vacuum Main Page, go to Object chamber | |
- | + | * Make sure that the IG is off. | |
- | ===== Pumping | + | * When you start, you should have RV closed, FV open and the TGV valve, Vent valve, as well as the gate valves named D2341 BGV and D2303 BGV all closed. |
- | Requirements: | + | * Close FV and open RV to start pumping down the object chamber with the roughing pump. You can make a plot of the pressure vs time at BPG at this point. |
- | * Verify | + | * Section D2337 will initially be red with vacuum status: 'Bad vacuum' |
- | * Verify that OBJ SCI is not biased by checking the HV module in dU6 | + | * Wait until the vacuum reading at FPG to drops to below 200 mTorr before you can safely turn on the turbo pump. |
- | * Open **S800Vac** | + | * After turning on the turbo pump, wait for the inner circle of the pump symbol to turn green and the pump to reach 40 000 rpm before you open the valve at TGV. |
- | * Check " | + | * When the pressure |
- | * When the pressure in **I191PG** is below 100 mTorr, the Turbo Pump will start automatically through | + | {{ : |
- | * **I190RV** will be closed | + | |
- | * **I190FV** will be opened | + | |
- | * **I188GV** will be opened | + | |
- | * When the pressure | + | |
- | + | ||