Line 1: Line 1:
 +====== Look Up table 4/3/2013-- ======
 +
 +^  Spectcl ​ ^  CRY id  ^  Qp  ^
 +^  00  |  26  |  Q2p3  |
 +^  01  |  30  |  Q6p3  |
 +^  02  |  25  |  Q2p2  |
 +^  03  |  29  |  Q6p2  |
 +^  04  |  27  |  Q2p4  |
 +^  05  |  31  |  Q6p4  |
 +^  06  |  24  |  Q2p1  |
 +^  07  |  28  |  Q6p1  |
 +^  08  |  47  |  Q5p4  |
 +^  09  |  51  |  Q4p4  |
 +^  10  |  59  |  Q3p4  |
 +^  11  |  63  |  Q1p4  |
 +^  12  |  44  |  Q5p1  |
 +^  13  |  50  |  Q4p3  |
 +^  14  |  58  |  Q3p3  |
 +^  15  |  60  |  Q1p1  |
 +^  16  |  46  |  Q5p3  |
 +^  17  |  48  |  Q4p1  |
 +^  18  |  56  |  Q3p1  |
 +^  19  |  62  |  Q1p3  |
 +^  20  |  45  |  Q5p2  |
 +^  21  |  49  |  Q4p2  |
 +^  22  |  57  |  Q3p2  |
 +^  23  |  61  |  Q1p2  |
 +^  24  |  88  |  Q7p1  |
 +^  25  |  91  |  Q7p4  |
 +^  26  |  89  |  Q7p2  |
 +^  27  |  90  |  Q7p3  |
 +
 +====== Look Up table 7/​19/​2012--4/​1/​2013 ========
 +
 +^  Spectcl ​ ^  CRY id  ^  Qp  ^ 
 +^  00  | 26 | ​ Q2p3  |
 +^  01  |  30  |  Q6p3  |
 +^  02  |  34  |  Q4p3  |
 +^  03  |  38  |  Q5p3  |
 +^  04  |  25  |  Q2p2  |
 +^  05  |  29  |  Q6p2  |
 +^  06  |  33  |  Q4p2  |
 +^  07  |  37  |  Q5p2  |
 +^  08  |  27  |  Q2p4  |
 +^  09  |  31  |  Q6p4  |
 +^  10  |  35  |  Q4p4  |
 +^  11  |  39  |  Q5p4  |
 +^  12  |  24  |  Q2p1  |
 +^  13  |  28  |  Q6p1  |
 +^  14  |  32  |  Q4p1  |
 +^  15  |  36  |  Q5p1  |
 +^  16  |  63  |  Q1p4  |
 +^  17  |  67  |  Q3p4  |
 +^  18  |  71  |  Q7p4  |
 +^  19  |  60  |  Q1p1  |
 +^  20  |  66  |  Q3p3  |
 +^  21  |  68  |  Q7p1  |
 +^  22  |  62  |  Q1p3  |
 +^  23  |  64  |  Q3p1  |
 +^  24  |  70  |  Q7p3  |
 +^  25  |  61  |  Q1p2  |
 +^  26  |  65  |  Q3p2  |
 +^  27  |  69  |  Q7p2  |
 +
 +
 GRETINA DAQ system GRETINA DAQ system
  
Line 99: Line 164:
   * CS_DataId in https://​wikihost.nscl.msu.edu/​gretina/​doku.php/​daq:​decomp   * CS_DataId in https://​wikihost.nscl.msu.edu/​gretina/​doku.php/​daq:​decomp
   * CS_GblId in /​global/​devel/​boot/​vmeArray/​vme*.cmd --- The Id is given by  hole_number * 16 + position_in_quad * 4 + board_number   * CS_GblId in /​global/​devel/​boot/​vmeArray/​vme*.cmd --- The Id is given by  hole_number * 16 + position_in_quad * 4 + board_number
-  * GTSort_shack/​DetNum.dat --- self explanatory+  * (GTSort_shack/​DetNum.dat --- self explanatory) <-- only if GTSort_shack is still in use for online spectra. 
 +  * Changes need to be made in non-linearity correction code from Heather (due to hardcoded detector positions). The file to be edited is in <​DAQversion#>/​gretClust/​gretClustApp/​src/​DecompIF.C ​  In this file, partway through that are a number of defines for the hole numbers for Q1-Q7. ​ These need to be changed to reflect the correct hole positions. ​ Then the DAQ codes needs to be recompiled (as user gretina, on dogs.gam), in the directory with the file, and two levels up (in the gretClust directory). ​ Ask Mario/​Heather for help if necessary. Problem will show up as RED Run Status on node display window. (Reason: ID in the code comes from hole#, if that is changed, wrong ID is calculated --> SegFault).
  
 ====== How to replace a broken node with a spare one ====== ====== How to replace a broken node with a spare one ======
Line 116: Line 182:
   * Switch on all the other vme crates   * Switch on all the other vme crates
   * wait till the green blinking lights appears on all the digitizers (~ 3 minutes)   * wait till the green blinking lights appears on all the digitizers (~ 3 minutes)
-  * start the daq and check the synchronization (alarm window or TS window) +  * start the daq and check the synchronization (alarm window or TS window). If few of the digitizers are out of sync recover them using the "​reboot IOCs button",​ otherwise power-cycle on their crate
-  * if few of the digitizers are dead recover them using the "​reboot IOCs button",​ otherwise power-cycle on their crate+
  