-====== Severity ====== 
-Scientific potential is reduced if 10MeV range doesn'​t provide good linearity in energy. ​ 
  
-====== Problem ====== 
-DNL of digitizers screw linearity, causing residue of >3keV for calibration function. 
- 
-====== Progress ====== 
-  
-5/18/12: Before the electronics were turned off at Berkeley, we completed the required pulser measurements to extract absolute non-linearity measurements for all 1120 ADC channels. ​ At this point, conversion from raw trace data to a useful non-linearity measurement is complete. ​ David Radford'​s correction code, is being implemented the correction in offline Mode3 data-sorting code.  Once successfully implemented,​ will discuss with Mario/Carl re: implementing in the pre-processor.  ​ 