-====== Severity ====== 
-Scientific potential is reduced if 10MeV range doesn'​t provide good linearity in energy. ​ 
- 
-====== Problem ====== 
-DNL of digitizers screw linearity, causing residue of >3keV for calibration function. 
- 
-====== Progress ====== 
-  
-5/18/12: Before the electronics were turned off at Berkeley, we completed the required pulser measurements to extract absolute non-linearity measurements for all 1120 ADC channels. ​ At this point, conversion from raw trace data to a useful non-linearity measurement is complete. ​ David Radford'​s correction code, is being implemented the correction in offline Mode3 data-sorting code.  Once successfully implemented,​ will discuss with Mario/Carl re: implementing in the pre-processor.  ​ 
- 
-5/29/12: David'​s correction code, which takes into account the 'local baseline',​ the resting baseline (equivalently the slope the pulse is sitting on) and the flat-top, has been implemented in offline sorting, and shows an improvement in the residuals from rms = 0.7keV to rms = 0.24keV (see plot), for old (CR-5) 152Eu calibration data.  The correction on the 10MeV scale channels appears to be working, at least at low rate, where the local and resting baselines are very close. ​ The high rate correction still needs to be tested for the 10MeV scale, which will require taking the appropriate data.  If possible, once I (Heather) arrive at NSCL, we can take the appropriate data and test the correction. ​ The next step for correcting 10MeV channels would then be to incorporate into pre-processing code (needs Mario'​s help). 
- 
-Corrections to the other gain CC channels has not yet been successful. ​ The measured INL corrections do not show agreement with the residuals as was the case for the 10MeV full scale data, which may indicate a problem with the extraction of the INL -- Heather and David will continue to look into this, and try to confirm the source of the problem. ​ It is possible that correction to the 30MeV CC may not be possible without more INL measurements. 
- 
- 
  

QR Code