 **Usual alarm types and response:** **Usual alarm types and response:**
  
-You need to log on gln1, see [[LN computer]].+You need to log on gln1, see [[LN computer]]. ​Go to directory ~/scripts and run ./​run-gln-ui.sh which gives the fill system interface.  
 +Type help for getting overview on commands. 
 + 
 +In some cases you need to edit the fill-system configuration file ~/​gln/​gln.fig for setting new temperature thresholds. In order 
 +to apply new gln.fig configuration you need to say '​load'​ in run-gln-ui.sh and MUST RE-ENABLE the filling! 
  
 __Long fill__ __Long fill__
 +  * What happened: The LN outlet sensor hasn't triggered. ​
 +  * Run run-gln-ui.sh
 +  * Say '​clr'​ to clear the alarm  ​
 +  * Check 'Fill Duration Plot' if this sensor got cold
 +  * If this happens regularly you may want to increase its trigger threshold in gln.fig (F1 directive, beginning at gln.fig)
 +
 +__Detector warm__
 +  * You get a detector high temperature alarm.
 +  * Check 'all zoom cold (pdf)' plot. 
 +  * T-plot for detector is just straight: Go to Detector warm I (spike)
 +  * T-plot for detector shows exp-like increase: Go to Detector warm II
 +
 +__Detector warm I__
 +  * What happened: T read spiked and caused T high alarm 
 +  * Run run-gln-ui.sh
 +  * Say '​clr'​ to clear alarm
 +  * You need to re-enable filling of this detector
 +
 + 
 + 