**Usual alarm types and response:** | **Usual alarm types and response:** | ||
- | You need to log on gln1, see [[LN computer]]. | + | You need to log on gln1, see [[LN computer]]. Go to directory ~/scripts and run ./run-gln-ui.sh which gives the fill system interface. |
+ | Type help for getting overview on commands. | ||
+ | |||
+ | In some cases you need to edit the fill-system configuration file ~/gln/gln.fig for setting new temperature thresholds. In order | ||
+ | to apply new gln.fig configuration you need to say 'load' in run-gln-ui.sh and MUST RE-ENABLE the filling! | ||
__Long fill__ | __Long fill__ | ||
+ | * What happened: The LN outlet sensor hasn't triggered. | ||
+ | * Run run-gln-ui.sh | ||
+ | * Say 'clr' to clear the alarm | ||
+ | * Check 'Fill Duration Plot' if this sensor got cold | ||
+ | * If this happens regularly you may want to increase its trigger threshold in gln.fig (F1 directive, beginning at gln.fig) | ||
+ | |||
+ | __Detector warm__ | ||
+ | * You get a detector high temperature alarm. | ||
+ | * Check 'all zoom cold (pdf)' plot. | ||
+ | * T-plot for detector is just straight: Go to Detector warm I (spike) | ||
+ | * T-plot for detector shows exp-like increase: Go to Detector warm II | ||
+ | |||
+ | __Detector warm I__ | ||
+ | * What happened: T read spiked and caused T high alarm | ||
+ | * Run run-gln-ui.sh | ||
+ | * Say 'clr' to clear alarm | ||
+ | * You need to re-enable filling of this detector | ||
+ | |||
+ | |||
+ |