 **Usual alarm types and response:** **Usual alarm types and response:**
  
-You need to log on gln1, see [[LN computer]].+You need to log on gln1, see [[LN computer]]. Go to directory ~/scripts and run ./​run-gln-ui.sh which gives the fill system interface.  
 +Type help for getting overview on commands. Get familiar with the commands mon, clr, ton, fill, coff, con, abort, load, quit 
 + 
 +In some cases you need to edit the fill-system configuration file ~/​scripts/​gln.fig for setting new temperature thresholds. In order 
 +to apply new gln.fig configuration you need to say '​load'​ in run-gln-ui.sh and MUST RE-ENABLE the filling! You can't say '​load'​  
 +while the system is filling. Sometimes you also need to exit and start the UI again before it allows you to execute '​load'​.  
 + 
 +IN WORST CASE: You can always hand-fill a quad in any mounting position.
  
 __Long fill__ __Long fill__
 +  * What happened: The LN outlet sensor hasn't triggered. ​
 +  * Run run-gln-ui.sh
 +  * Say '​clr'​ to clear the alarm  ​
 +  * Check 'Fill Temperature Plot' if this sensor got cold
 +  * If this happens regularly you may want to increase its trigger threshold in gln.fig (F1 directive, beginning in gln.fig)
 +
 +__Detector warm__
 +  * You get a detector high temperature alarm.
 +  * Check 'all zoom cold (pdf)' plot. 
 +  * T-plot for detector is just straight: Go to Detector warm I (spike)
 +  * T-plot for detector shows exp-like increase: Go to Detector warm II
 +
 +__Detector warm I (spike)__
 +  * What happened: T read spiked and caused T high alarm 
 +  * Run run-gln-ui.sh
 +  * Say '​clr'​ to clear alarm
 +  * You need to re-enable filling of this detector (ton #det)
 +  * If this happens regularly you may want to increase the T high threshold in gln.fig (F3 directive, end in gln.fig)
 +
 +__Detector warm II__
 +  * What happened: Detector runs out of LN and really warms up
 +  * IMPORTANT: detector T increased by few Kelvin only, see example plot below.
 +  * Run run-gln-ui.sh
 +  * Say 'fill det#', confirm that you really want to fill it.
 +  * Say 'coff det#', this stops T checking (and alarming and disabling) of this detector
 +  * Watch the fill, check for reasonable fill time (not that LN out is triggering much too early)
 +  * After fill, within ~30 min the temperature should recover (see example plot below).
 +  * If T is good say 'con #det', so T checking is enabled again.
 +  * If it is deep in the night: You can temporary increase T high threshold (F3 directive)
 +  * NEVER LEAVE DETECTOR T CHECKED OFF 
 + 
 +
 +
 +Here an real-life example. After alarm got received you get following T-plot:
 +{{:​fillsystem:​warm-up-example.pdf|}}. Actually after alarm you get it up to 
 +channel ~1100 and see T is going up. Therefore it is a real alarm. Next you do
 +is logging on the gln system, run the UI, and say 'fill 6 now'. Then clr and directly
 +coff 6 (as a fill enables T checking by default). Later you see how T is recovering.
 +dirk cell:517 203 9419
 +
 +
 + 