Line 7: | Line 7: | ||
to apply new gln.fig configuration you need to say 'load' in run-gln-ui.sh and MUST RE-ENABLE the filling! | to apply new gln.fig configuration you need to say 'load' in run-gln-ui.sh and MUST RE-ENABLE the filling! | ||
+ | IN WORST CASE: You can always hand-fill a quad in any mounting position. | ||
__Long fill__ | __Long fill__ | ||
Line 13: | Line 14: | ||
* Say 'clr' to clear the alarm | * Say 'clr' to clear the alarm | ||
* Check 'Fill Duration Plot' if this sensor got cold | * Check 'Fill Duration Plot' if this sensor got cold | ||
- | * If this happens regularly you may want to increase its trigger threshold in gln.fig (F1 directive, beginning at gln.fig) | + | * If this happens regularly you may want to increase its trigger threshold in gln.fig (F1 directive, beginning in gln.fig) |
__Detector warm__ | __Detector warm__ | ||
Line 21: | Line 22: | ||
* T-plot for detector shows exp-like increase: Go to Detector warm II | * T-plot for detector shows exp-like increase: Go to Detector warm II | ||
- | __Detector warm I__ | + | __Detector warm I (spike)__ |
* What happened: T read spiked and caused T high alarm | * What happened: T read spiked and caused T high alarm | ||
* Run run-gln-ui.sh | * Run run-gln-ui.sh | ||
* Say 'clr' to clear alarm | * Say 'clr' to clear alarm | ||
* You need to re-enable filling of this detector | * You need to re-enable filling of this detector | ||
+ | * If this happens regularly you may want to increase the T high threshold in gln.fig (F3 directive, end in gln.fig) | ||
+ | __Detector warm II__ | ||
+ | * What happened: Detector runs out of LN and really warms up | ||
+ | * IMPORTANT: detector T increased by few Kelvin only, see example plot. | ||
+ | * Run run-gln-ui.sh | ||
+ | * Say 'fill det#', confirm that you want to fill it. | ||
+ | * Say 'coff det#', this stops T checking (and alarming and disabling) of this detector | ||
+ | * Watch the fill, check for reasonable fill time (not that LN out is triggering much too early) | ||
+ | * After fill, within ~30 min the temperature should recover (see example plot). | ||
+ | * If T is good say 'con #det', so T checking is enabled again. | ||
+ | * If it is deep in the night: You can temporary increase T high threshold (F3 directive) | ||
+ | * NEVER LEAVE DETECTOR T CHECK OFF | ||
- | + | ||
+ | {{:fillsystem:warm-up-example.pdf|}} |