Line 7: Line 7:
 to apply new gln.fig configuration you need to say '​load'​ in run-gln-ui.sh and MUST RE-ENABLE the filling! to apply new gln.fig configuration you need to say '​load'​ in run-gln-ui.sh and MUST RE-ENABLE the filling!
  
 +IN WORST CASE: You can always hand-fill a quad in any mounting position.
  
 __Long fill__ __Long fill__
Line 13: Line 14:
   * Say '​clr'​ to clear the alarm  ​   * Say '​clr'​ to clear the alarm  ​
   * Check 'Fill Duration Plot' if this sensor got cold   * Check 'Fill Duration Plot' if this sensor got cold
-  * If this happens regularly you may want to increase its trigger threshold in gln.fig (F1 directive, beginning ​at gln.fig)+  * If this happens regularly you may want to increase its trigger threshold in gln.fig (F1 directive, beginning ​in gln.fig)
  
 __Detector warm__ __Detector warm__
Line 21: Line 22:
   * T-plot for detector shows exp-like increase: Go to Detector warm II   * T-plot for detector shows exp-like increase: Go to Detector warm II
  
-__Detector warm I__+__Detector warm I (spike)__
   * What happened: T read spiked and caused T high alarm    * What happened: T read spiked and caused T high alarm 
   * Run run-gln-ui.sh   * Run run-gln-ui.sh
   * Say '​clr'​ to clear alarm   * Say '​clr'​ to clear alarm
   * You need to re-enable filling of this detector   * You need to re-enable filling of this detector
 +  * If this happens regularly you may want to increase the T high threshold in gln.fig (F3 directive, end in gln.fig)
  
 +__Detector warm II__
 +  * What happened: Detector runs out of LN and really warms up
 +  * IMPORTANT: detector T increased by few Kelvin only, see example plot.
 +  * Run run-gln-ui.sh
 +  * Say 'fill det#', confirm that you want to fill it.
 +  * Say 'coff det#', this stops T checking (and alarming and disabling) of this detector
 +  * Watch the fill, check for reasonable fill time (not that LN out is triggering much too early)
 +  * After fill, within ~30 min the temperature should recover (see example plot).
 +  * If T is good say 'con #det', so T checking is enabled again.
 +  * If it is deep in the night: You can temporary increase T high threshold (F3 directive)
 +  * NEVER LEAVE DETECTOR T CHECK OFF 
    
- + 
 +{{:​fillsystem:​warm-up-example.pdf|}}