Line 2: Line 2:
  
 You need to log on gln1, see [[LN computer]]. Go to directory ~/scripts and run ./​run-gln-ui.sh which gives the fill system interface. ​ You need to log on gln1, see [[LN computer]]. Go to directory ~/scripts and run ./​run-gln-ui.sh which gives the fill system interface. ​
-Type help for getting overview on commands. Get familiar with the commands ​fill, clr, coff, con, abort, load, quit+Type help for getting overview on commands. Get familiar with the commands ​mon, clr, ton, fill, coff, con, abort, load, quit
  
 In some cases you need to edit the fill-system configuration file ~/​gln/​gln.fig for setting new temperature thresholds. In order In some cases you need to edit the fill-system configuration file ~/​gln/​gln.fig for setting new temperature thresholds. In order
Line 27: Line 27:
   * Run run-gln-ui.sh   * Run run-gln-ui.sh
   * Say '​clr'​ to clear alarm   * Say '​clr'​ to clear alarm
-  * You need to re-enable filling of this detector+  * You need to re-enable filling of this detector ​(ton #det)
   * If this happens regularly you may want to increase the T high threshold in gln.fig (F3 directive, end in gln.fig)   * If this happens regularly you may want to increase the T high threshold in gln.fig (F3 directive, end in gln.fig)
  
Line 34: Line 34:
   * IMPORTANT: detector T increased by few Kelvin only, see example plot below.   * IMPORTANT: detector T increased by few Kelvin only, see example plot below.
   * Run run-gln-ui.sh   * Run run-gln-ui.sh
-  * Say 'fill det#', confirm that you want to fill it.+  * Say 'fill det#', confirm that you really ​want to fill it.
   * Say 'coff det#', this stops T checking (and alarming and disabling) of this detector   * Say 'coff det#', this stops T checking (and alarming and disabling) of this detector
   * Watch the fill, check for reasonable fill time (not that LN out is triggering much too early)   * Watch the fill, check for reasonable fill time (not that LN out is triggering much too early)
Line 40: Line 40:
   * If T is good say 'con #det', so T checking is enabled again.   * If T is good say 'con #det', so T checking is enabled again.
   * If it is deep in the night: You can temporary increase T high threshold (F3 directive)   * If it is deep in the night: You can temporary increase T high threshold (F3 directive)
-  * NEVER LEAVE DETECTOR T CHECK OFF +  * NEVER LEAVE DETECTOR T CHECKED ​OFF 
    
  
Line 47: Line 47:
 {{:​fillsystem:​warm-up-example.pdf|}}. Actually after alarm you get it up to  {{:​fillsystem:​warm-up-example.pdf|}}. Actually after alarm you get it up to 
 channel ~1100 and see T is going up. Therefore it is a real alarm. Next you do channel ~1100 and see T is going up. Therefore it is a real alarm. Next you do
-is logging on the gln system, run the UI, and say fill 6. Then clr and directly+is logging on the gln system, run the UI, and say 'fill 6 now'. Then clr and directly
 coff 6 (as a fill enables T checking by default). Later you see how T is recovering. coff 6 (as a fill enables T checking by default). Later you see how T is recovering.
 +dirk cell:517 203 9419
  
  
    