fillsystem:alarmresponse [2013/05/23 08:53]
weisshaa
fillsystem:alarmresponse [2014/02/27 16:14]
127.0.0.1 external edit
Line 34: Line 34:
   * IMPORTANT: detector T increased by few Kelvin only, see example plot below.   * IMPORTANT: detector T increased by few Kelvin only, see example plot below.
   * Run run-gln-ui.sh   * Run run-gln-ui.sh
-  * Say 'fill det# now', confirm that you want to fill it.+  * Say 'fill det#', confirm that you really ​want to fill it.
   * Say 'coff det#', this stops T checking (and alarming and disabling) of this detector   * Say 'coff det#', this stops T checking (and alarming and disabling) of this detector
   * Watch the fill, check for reasonable fill time (not that LN out is triggering much too early)   * Watch the fill, check for reasonable fill time (not that LN out is triggering much too early)
Line 40: Line 40:
   * If T is good say 'con #det', so T checking is enabled again.   * If T is good say 'con #det', so T checking is enabled again.
   * If it is deep in the night: You can temporary increase T high threshold (F3 directive)   * If it is deep in the night: You can temporary increase T high threshold (F3 directive)
-  * NEVER LEAVE DETECTOR T CHECK OFF +  * NEVER LEAVE DETECTOR T CHECKED ​OFF 
    
  

QR Code