Line 4: Line 4:
 Type help for getting overview on commands. Get familiar with the commands mon, clr, ton, fill, coff, con, abort, load, quit Type help for getting overview on commands. Get familiar with the commands mon, clr, ton, fill, coff, con, abort, load, quit
  
-In some cases you need to edit the fill-system configuration file ~/gln/gln.fig for setting new temperature thresholds. In order+In some cases you need to edit the fill-system configuration file ~/scripts/gln.fig for setting new temperature thresholds. In order
 to apply new gln.fig configuration you need to say '​load'​ in run-gln-ui.sh and MUST RE-ENABLE the filling! You can't say '​load' ​ to apply new gln.fig configuration you need to say '​load'​ in run-gln-ui.sh and MUST RE-ENABLE the filling! You can't say '​load' ​
 while the system is filling. Sometimes you also need to exit and start the UI again before it allows you to execute '​load'​. ​ while the system is filling. Sometimes you also need to exit and start the UI again before it allows you to execute '​load'​. ​
Line 14: Line 14:
   * Run run-gln-ui.sh   * Run run-gln-ui.sh
   * Say '​clr'​ to clear the alarm  ​   * Say '​clr'​ to clear the alarm  ​
-  * Check '​Fill ​Duration ​Plot' if this sensor got cold+  * Check '​Fill ​Temperature ​Plot' if this sensor got cold
   * If this happens regularly you may want to increase its trigger threshold in gln.fig (F1 directive, beginning in gln.fig)   * If this happens regularly you may want to increase its trigger threshold in gln.fig (F1 directive, beginning in gln.fig)
  
Line 32: Line 32:
 __Detector warm II__ __Detector warm II__
   * What happened: Detector runs out of LN and really warms up   * What happened: Detector runs out of LN and really warms up
-  * IMPORTANT: detector T increased by few Kelvin only, see example plot below.+  * IMPORTANT: detector T increased by few Kelvin only, see example plot {{:​fillsystem:​warm-up-example.pdf|}}.
   * Run run-gln-ui.sh   * Run run-gln-ui.sh
   * Say 'fill det#', confirm that you really want to fill it.   * Say 'fill det#', confirm that you really want to fill it.