Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Last revision Both sides next revision
fillsystem:ln_computer [2013/05/07 11:08]
weisshaa
fillsystem:ln_computer [2019/05/17 12:32]
weisshaa [Login on LN computer]
Line 1: Line 1:
 ====== Login on LN computer ====== ====== Login on LN computer ======
-The LN computer lives in the GRETINA net and doesn'​t know +The LN computer lives in the GRETINA net and since December 2015 it is accessible from NSCL network
-about the outside world.+
  
   * login on gateway machine nsclgw1.nscl.msu.edu   * login on gateway machine nsclgw1.nscl.msu.edu
-  * login as gretina ​on a2.gam (35.8.35.163+  * login as ln on gretina-gln1.nscl.msu.edu ​(35.8.35.167
-  * login as ln on gln1 (10.0.4.11) +  * Remember: "​liquid nitrogen two for gretina"​ 
-  * Rememeber: "​liquid nitrogen two for gretina"​+  * from a1/a2: ln@gln1
  
 ====== LN scripts ====== ====== LN scripts ======
Line 27: Line 26:
   *run-mail-que-flush.sh  ​   *run-mail-que-flush.sh  ​
   *run-view-master.sh:​ Output of the LN master process   *run-view-master.sh:​ Output of the LN master process
 +
 +Making long term T plots:
 +
 +  *on gln machine go to ~/​other/​thistory-dev
 +  *run there ./​run-thistory.sh XXX with the last day as input parameter
 +  *results th.pdf and d.pdf are written in ~/plots
 +  *uncomment/​comment different templates in script for covering different time ranges
 +
  
 ====== LN configuration ====== ====== LN configuration ======
-The LN configuration file is found as ~/​gln/​gln.fig. It is //​self-explanatory//​. If this file got updated, use run-gln-ui.sh script and the command "​load"​ in there. Note: After a "​load"​ in the UI all detector fill schedules are gone! You need to do a "fill x y z after h:​mm"​. ​  +The LN configuration file is found as ~/​gln/​gln.fig. It is //​self-explanatory//​. If this file got updated, use run-gln-ui.sh script and the command "​load"​ in there. Note: After a "​load"​ in the UI all detector fill schedules are gone! You need to do a "ton x y z ..." or "fill x y z after h:​mm"​. ​  
  
 ====== ADC to channel mapping ====== ====== ADC to channel mapping ======
-Documentation for ADC mapping: {{:​connections_1_.pdf|}}+Documentation for ADC mapping: {{:​connections_1_.pdf|}} ​(OUTDATED, now we have silver boxes)
  
   * ADC 1-4: Manifold overflow 1-4 (Manifold overflow at supply box Q1-Q4)   * ADC 1-4: Manifold overflow 1-4 (Manifold overflow at supply box Q1-Q4)

QR Code
QR Code fillsystem:ln_computer (generated for current page)