Line 112: Line 112:
 Decreasing these parameters only increases the mass resolution. It seems the smaller the beam spot, the larger the mass resolution. Decreasing these parameters only increases the mass resolution. It seems the smaller the beam spot, the larger the mass resolution.
  
-b) Adjusting the pitch, roll, yaw, x-position, and y-position of Q2 caused a drop in mass resolution at the mass slits (FP2), right after the Wien filter.  In COSY, "TA" is used to adjust the pitch and yaw, "RA" to adjust the roll, and "SA" to adjust x and y position.  Since the shifts are made relative to the central ray, they must be "undone" directly after the quadruple to prevent the shift form effecting the downstream elements.  The below plots show the relative effects on M_res.  For Q2 the mass resolution is much more sensitive to changes in pitch and x-offset.  +b) Adjusting the pitch, roll, yaw, x-position, and y-position of Q2 caused a drop in mass resolution at the mass slits (FP2), right after the Wien filter.  In COSY, "TA" is used to adjust the pitch and yaw, "RA" to adjust the roll, and "SA" to adjust x and y position.  Since the shifts are made relative to the central ray, they must be "undone" directly after the quadruple to prevent the shift form effecting the downstream elements.  The below plots show the relative effects on M_res.  For Q2 the mass resolution is more sensitive to changes in pitch and x-offset.   
 + 
 +{{ :angular_shifts.png |}} 
 +{{ :position_shifts.png |}} 
 + 
/srv/thewikis/JIOSS/data/attic/group_10.1536931085.txt.gz · Last modified: 2018/09/14 09:18 by seymour