Line 31: Line 31:
 __Detector warm II__ __Detector warm II__
   * What happened: Detector runs out of LN and really warms up   * What happened: Detector runs out of LN and really warms up
-  * IMPORTANT: detector T increased by few Kelvin only, see example plot.+  * IMPORTANT: detector T increased by few Kelvin only, see example plot below.
   * Run run-gln-ui.sh   * Run run-gln-ui.sh
   * Say 'fill det#', confirm that you want to fill it.   * Say 'fill det#', confirm that you want to fill it.
   * Say 'coff det#', this stops T checking (and alarming and disabling) of this detector   * Say 'coff det#', this stops T checking (and alarming and disabling) of this detector
   * Watch the fill, check for reasonable fill time (not that LN out is triggering much too early)   * Watch the fill, check for reasonable fill time (not that LN out is triggering much too early)
-  * After fill, within ~30 min the temperature should recover (see example plot).+  * After fill, within ~30 min the temperature should recover (see example plot below).
   * If T is good say 'con #det', so T checking is enabled again.   * If T is good say 'con #det', so T checking is enabled again.
   * If it is deep in the night: You can temporary increase T high threshold (F3 directive)   * If it is deep in the night: You can temporary increase T high threshold (F3 directive)
Line 42: Line 42:
    
  
 +
 +Here an real-life example. After alarm got received ​
 {{:​fillsystem:​warm-up-example.pdf|}} {{:​fillsystem:​warm-up-example.pdf|}}